ABSTRACT In this prospective, multi-reader, multi-vendor study, we evaluated the performance of a commercially available deep neural network (DNN)–based MR image reconstruction in enabling
accelerated 2D fast spin-echo (FSE) knee imaging. Forty-five subjects were prospectively enrolled and randomly divided into three 3T MRIs. Conventional 2D FSE and accelerated 2D FSE
sequences were acquired for each subject, and the accelerated FSE images were reconstructed and enhanced with DNN–based reconstruction software (FSE-DNN). Quantitative assessments and
diagnostic performances were independently evaluated by three musculoskeletal radiologists. For statistical analyses, paired _t_-tests, and Pearson’s correlation were used for image quality
comparison and inter-reader agreements. Accelerated FSE-DNN reduced scan times by 41.0% on average. FSE-DNN showed better SNR and CNR (_p_ < 0.001). Overall image quality of FSE-DNN was
comparable (_p_ > 0.05), and diagnostic performances of FSE-DNN showed comparable lesion detection. Two of cartilage lesions were under-graded or over-graded (n = 2) while there was no
significant difference in other image sets (n = 43). Overall inter-reader agreement between FSE-conventional and FSE-DNN showed good agreement (R2 = 0.76; _p_ < 0.001). In conclusion,
DNN-based reconstruction can be applied to accelerated knee imaging in multi-vendor MRI scanners, with reduced scan time and comparable image quality. This study suggests the potential for
DNN-accelerated knee MRI in clinical practice. SIMILAR CONTENT BEING VIEWED BY OTHERS DIAGNOSTIC ADVANTAGE OF THIN SLICE 2D MRI AND MULTIPLANAR RECONSTRUCTION OF THE KNEE JOINT USING DEEP
LEARNING BASED DENOISING APPROACH Article Open access 20 June 2022 FASTMRI+, CLINICAL PATHOLOGY ANNOTATIONS FOR KNEE AND BRAIN FULLY SAMPLED MAGNETIC RESONANCE IMAGING DATA Article Open
access 05 April 2022 REGION OF INTEREST-SPECIFIC LOSS FUNCTIONS IMPROVE T2 QUANTIFICATION WITH ULTRAFAST T2 MAPPING MRI SEQUENCES IN KNEE, HIP AND LUMBAR SPINE Article Open access 23
December 2022 INTRODUCTION It is reported that about 25% of adults experience knee pain, resulting in limitations to their functional capabilities and mobility, and causing a negative impact
on their quality of life, and the prevalence of knee pain has shown an upward trend over time, regardless of age1. Magnetic resonance imaging (MRI) plays an important role in evaluating
internal derangements in patients with knee pain2. Common indications of knee MRI are trauma, overuse, degeneration, and knee pain. Commonly used knee MRI protocols are parallel image based
MR sequences with multi-channel phased array coil3,4. Typical sequences are triplane fat-suppressed fluid-sensitive, sagittal PD-weighted, and coronal or axial T1-weighted image with 15–25
min scan time. 3D sequence or deep learning reconstructions may be added5. Common MRI protocols consisting of four to five separately acquired 2D fast spin-echo (FSE) or 2D turbo spin-echo
(TSE) pulse sequences are commonly used as a standard in clinical practice, providing excellent tissue contrast and high spatial resolution, enabling good assessment of meniscal,
ligamentous, and cartilaginous injuries6. Accurate and noninvasive imaging evaluation requires three planes of axial, coronal, and sagittal fat-saturated proton density-weighted or
intermediate-weighted images, which require repetitive scans with relatively long scan times. Accelerated MRI is essential in knee imaging because patients with knee pain tend to move,
causing motion artifacts, especially when scan time is prolonged. Long scan time of MRI scans can result in reduced productivity per MRI scanner and elevated MRI cost7. Implementing
accelerated MRI techniques can alleviate patient discomfort and enhance the cost-effectiveness of the process. Recent advances in various accelerated imaging methods have shown the
feasibility of accelerated knee MRI, in some cases enabling a 5-min knee imaging protocol8,9,10. Parallel imaging (PI) is one approach to accelerate MRI data acquisition, and it is based on
the principle of acquiring spatial encoding data from overlapping phased-array coil elements that sample the MR signal in parallel11. Although disadvantages of PI include reduced
signal-to-noise ratio (SNR), aliasing, and reconstruction-related artifacts, acceleration with PI allows for rapid imaging due to the advancement of multi-channel phased-array coil
technology12. In the knee, 2D FSE with PI has been widely utilized for routine 2D FSE protocols9. However, PI acceleration factors higher than 2 cannot be reliably achieved in clinical
settings without compromising image quality13. Compressed sensing (CS) was developed on the premise of reconstructing an image from an under-sampled _k_-space, since the number of data
segments in the _k_-space is a direct determinant of image acquisition time14. The combination of CS and PI allows even faster imaging, with the resultant image quality deemed acceptable15.
However, they require a high computational burden during the image reconstruction process with long iteration times, limiting their use in routine clinical practice. Recently, deep neural
network (DNN)–based MRI reconstructions have been proposed, showing great potential to reduce MRI acquisition time16,17. Deep learning-based MRI reconstruction techniques have been approved
and are being evaluated in clinical practice18. Currently, the software requires to be evaluated and monitored from its premarket development to post-market performance in real-world
radiology19. However, to date, there has been only a few multi-vendor studies20,21,22 that evaluated the image quality and performances with commercially available DNN–based magnetic
resonance imaging (MRI) reconstruction. The purpose of this prospective, multi-reader, multi-vendor study was to evaluate the performance of commercially available DNN-based MR image
reconstruction software in enabling accelerated 2D FSE knee imaging in a clinical environment. We hypothesized that highly accelerated 2D FSE knee imaging combined with DNN–based
reconstruction would allow a decrease in scan time while yielding comparable image quality and diagnostic performance for ligamentous, meniscal, and cartilaginous lesions against
conventional 2D FSE knee MRI. MATERIALS AND METHODS AIRS Medical provided financial support for this prospective study. The authors had control of the data and the information submitted for
publication. STUDY POPULATION This prospective study from a single tertiary center was approved by the Institutional Review Board of Yonsei University’s Health System (IRB No: 1-2022-0017).
Written informed consent was obtained from all enrolled participants. Our study complied with both the Declaration of Helsinki and the Health Insurance Portability and Accountability Act.
Study recruitment commenced from August 2022 to October 2022. Inclusion criteria were: (1) clinically indicated patient for knee MRI; (2) an agreement to participate in DNN accelerated knee
MRI; (3) age of 30 years or older; (4) the ability to position the knee in MRI; and (5) symptomatic knees associated with pain and dysfunction knee. Exclusion criteria consisted of (1)
orthopedic implants in the knee region and (2) other general contraindications for MRI. The flow chart for prospective study enrollment is shown in Fig. 1. MR IMAGING PROTOCOL All enrolled
patients were randomly assigned to MRI exams with three different 3.0 T MR scanners, and they underwent knee MRI with both conventional and DNN-accelerated MRI sequences: axial fat-saturated
T2-weighted image, coronal fat-saturated T2-weighted image, sagittal T2-weighted image, and axial T1-weighted image. MR imaging was performed on three 3.0 T MR scanners (Ellition X, Philips
Healthcare, Best, The Netherlands; Prisma Fit, Siemens Healthineers, Erlangen, Germany; Discovery MR750, GE Healthcare, Waukesha, WI, USA) with a dedicated 16-channel knee coil (Philips
Healthcare), a dedicated 16-channel knee coil (Siemens Healthineers), and a dedicated 8-channel HD transmit/receive knee coil (GE Healthcare), respectively. Conventional image acquisition
utilized knee imaging protocols routinely used at our institution, and the accelerated protocols were achieved by modifying scan time-related MRI parameters. Overall scan time reductions of
the DNN-accelerated MRI sequences were 41.0% (43.1%, Philips Healthcare; 38.1%, Siemens Healthineers, 41.72%, GE Healthcare). Accelerated 2D FSE utilized parallel imaging alone such as
Sensitivity Encoding (SENSE), Array coil Spatial Sensitivity Encoding (ASSET), and Generalized Autocalibrating Partial Parallel Acquisition (GRAPPA) depending on vendor’s technique.
Conventional 2D FSE images utilized combination of compressed sensing (CS) and parallel imaging such as CS-SENSE. A detailed summary of both acquisition methods of MRI parameters is shown in
Table 1. DEEP NEURAL NETWORK (DNN)–BASED IMAGE RECONSTRUCTION Commercially available deep neural network (DNN)–based MR image reconstruction software was used to reconstruct the accelerated
acquisition images (SwiftMR, v2.0.1.0. AIRS medical, Seoul, Korea). The software algorithm was based on the popular 2D U-net structure23 widely used in deep learning architectures in
various medical imaging applications. In this model, 18 convolutional blocks, 4 max-pooling layers, 4 up-sampling layers, 4 feature concatenations, and 3 convolutional layers were
incorporated in a cascade, with each layer enforcing data consistency. The model was trained and internally validated with 31,865 series and 3540 series of MR images, respectively. The model
underwent training using images from the entire body, considering also the musculoskeletal images including the knee. All imaging sequences with different contrasts commonly used in the
clinical practice were are included as well. Additionally, the model’s loss function was defined as the structural similarity index (SSIM) between the input and the label image, and the
model was optimized with Adam24 over 20 epochs using batch size of 4 at a learning rate of 10–3, decaying to 10–4. The network was trained using four NVIDIA Tesla V100 GPUs with 32 GB memory
(NVIDIA Corporation, CA, USA). to evaluate FSE-conventional and accelerated MR sequences with FSE-DNN. Schematic diagram of FSE-DNN architecture was in Fig. 2. The algorithm includes a deep
convolutional neural network (CNN) component that removes noise in the image domain and estimates the truncated high-frequency image data. This pipeline can be applied to 2D and 3D
acquisitions in multiple anatomic regions and for various pulse sequences, contrast weightings, field strengths, and coil configurations. The amount of noise reduction could be controlled,
owing to the model’s training process incorporating varying levels of noise amount on the input side. For this study, noise reduction level of low (51% reduction) was used because this level
had been found to yield the most similar perceived image quality when compared to the standard images. QUANTITATIVE IMAGE QUALITY ANALYSIS For a quantitative comparison of image quality,
signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated for all images acquired for this study (conventional, accelerated, and DNN-reconstructed). The femoral bone
marrow signal shown in the representative center slice in each image was calculated from a circular region of interest (ROI) placed in the same location for all images, and the mean standard
deviation (SD) of the background noise was used for SNR calculation. For CNR, three different imaging plane-landmark combinations were considered—(1) for axial T1-weighted and T2-weighted
fat saturated images, marrow-to-muscle signal difference (medial femoral condyle and biceps femoris muscle) and mean SD of background noise were used; (2) for coronal T2-weighted fat
saturated images, marrow-to-meniscus signal difference (medial femoral condyle and medial meniscus) and mean SD of background noise were used; (3) for sagittal T2-weighted images,
marrow-to-tendon signal difference(distal femur and patellar tendon) and mean SD of background noise were used to calculate the CNR. SUBJECTIVE IMAGE QUALITY ANALYSIS Three board-certified
musculoskeletal radiologists with six years (J.L.), one year (N.L.), and one year (Y.L.) of subspecialty experience individually assessed the conventional and DNN-reconstructed image sets of
the 45 knee MRI, using picture archiving and communication system (PACS) monitors (Totoku, Tokyo, Japan). The radiologists were blind to whether the image was FSE-conventional or FSE-DNN.
Each image was assigned a unique random number, and the images were evaluated in a random order. The reviewers scored the images independently based on clarity of overall image quality of
anatomical structures, perceived image noise, presence of imaging artifacts using a five-point scale. The clear visibility of the medial meniscus and lateral meniscus; ligamentous structures
including anterior cruciate ligament (ACL), posterior cruciate ligament (PCL), and medial collateral ligament (MCL); and cartilage grading were estimated separately. DIAGNOSTIC PERFORMANCE
FOR LIGAMENTOUS, MENISCAL, AND CARTILAGINOUS LESIONS The three radiologists (J.L., N.L., and Y.L.) formed a consensus on grading meniscal lesions, detecting ligamentous lesions, and grading
cartilage by reviewing MR images of non-enrolled patients together before evaluation. The radiologists were blinded to the medical records associated with the images acquired for this study.
Three radiologists independently evaluated the knee MRI studies for meniscal, tendinous, ligamentous, and osseocartilaginous injuries. For diagnostic performance analysis, all MRI findings
and clinical medical records were reviewed to form a reference standard by two additional reviewers (Y.H.L. and M.J.). The diagnostic criteria for meniscal tears were the presence of
abnormal signal intensity within the meniscus extending to the meniscal articular surface and abnormal morphology of the meniscus (0, normal; 1, tear)25,26. The diagnostic criteria for ACL
are grade 1, minimal lesion with T2 hyperintensities; grade 2, 50% or less injury with T2 hyperintensities; grade 3, complete injury (0, normal; 1, grade 1; 2, grade 2; 3, grade 3)27. The
diagnostic criteria for PCL tear are an abnormal T2 hyperintensity combined with a discontinuous appearance of the PCL and FCL fibers (0, normal; 1, tear)28,29. Intraligamentous ganglion
were recorded as grade 1 in ACL and PCL. The diagnostic criteria for MCL tear are grade 1, periligamentous T2 hyperintense with no fiber discontinuity; grade 2, partial discontinuity of the
fibers; grade 3, complete ligament (0, normal; 1, grade 1; 2, grade 2; 3, grade 3)29. For qualitative analyses of cartilage grading, the radiologists used the Outerbridge classification
system30,31 : cartilage grade, Grade 0 = intact cartilage; grade 1 = signal change on T2-weighted MR images; grade 2 = cartilage defect less than 50 percent of the depth; grade 3 = cartilage
defect 50% or more of the depth; and grade 4 = full-thickness cartilage defect with exposure of subchondral bone. When multiple cartilage lesions were present, the cartilage lesion with the
highest grade was recorded. STATISTICAL ANALYSIS Paired _t_-tests were performed to assess the statistical significance of the difference in the quantitative evaluation of SNR and CNR. For
the subjective analysis, we calculated the difference in the qualitative image quality score of anatomical structures, perceived image noise, presence of imaging artifacts by using a paired
_t_-test, and inter-reader agreement was assessed using Pearson’s correlation. Diagnostic performances of the FSE-DNN were analyzed in terms of sensitivity, specificity, area under curve
(AUC), and accuracy. To assess the diagnostic performance of the images in the cartilaginous lesion, the agreements of FSE-conventional and FSE-DNN were assessed using Pearson’s correlation.
All statistical analyses were performed in MedCalc (MedCalc Software, Ostend, Belgium) and Microsoft Excel (Microsoft, Redmond, WA, USA). _P_-values < 0.05 were considered statistically
significant. RESULTS DEMOGRAPHIC CHARACTERISTICS AND SCAN TIME REDUCTION Forty-five patients who underwent this research protocol of knee MRI including routine and accelerated MR pulse
sequence were enrolled in three vendors evenly (15 patients for each scanner). The age range of the 45 patients was 30–78 years (mean age ± standard deviation, 53.9 ± 11.8 years). Fourteen
patients were male and 31 were female. A total of 45 MRIs of three-vendors were evaluated. Accelerated FSE-DNN reduced scan times by average 41.0% compared to FSE-conventional (GE 41.7%,
Philips 43.1%, Siemens 38.1%), respectively. QUANTITATIVE IMAGE QUALITY ANALYSIS SNR and CNR on accelerated FSE images were significantly decreased with scan time reduction such as changing
parallel imaging factors and the number of phase encoding steps. FSE-DNN reconstruction software could enhance the SNR and CNR of accelerated FSE image in a short time. FSE-DNN showed
statistically better SNR and CNR than convention FSE : SNR ratio were 2.06, 2.23, 2.63, and 2.10 on axial T2-weighted fat-saturated FSE, axial T1-weighted FSE, sagittal T2-weighted FSE, and
coronal T2-weighted fat-saturated FSE, respectively ; CNR radio were 2.14, 3.00, 2.81, and 1.83 of marrow-to-muscle on axial T2-weighted fat-saturated FSE, marrow-to-muscle on axial
T1-weighted FSE, marrow-to-meniscus on sagittal T2-weighted FSE, and marrow-to-tendon on coronal T2-weighted fat-saturated FSE, respectively. Scan reductions were 44.63%, 45.56%, 29.80%, and
43.87% in axial T2-weighted fat-saturated FSE, axial T1-weighted FSE, sagittal T2-weighted FSE, and coronal T2-weighted fat-saturated FSE, respectively. The image quality analyses with
paired t-test results were summarized in Table 2. SUBJECTIVE IMAGE QUALITY ANALYSIS For qualitative evaluation, overall image quality of FSE-DNN was comparable (_p_ > 0.05), depending on
the reader. Overall image qualities of anatomical structures were better or equal (n = 42/45 in reviewer 1, n = 44/45 in reviewer 2, and n = 43/45 in reviewer 3). Image noise scores were
higher or equal (n = 42/45 in reviewer 1 and 2, n = 44/45 in reviewer 3). Imaging artifacts scores were higher or equal (n = 40/45 in reviewer 1, n = 41/45 in reviewer 2, and n = 42/45 in
reviewer 3). The qualitative evaluation of three radiologists were displayed in Fig. 3. The comparisons between FSE-conventional and FSE-DNN images were shown at overall image quality of
anatomical structures, perceived image noise, presence of imaging artifacts using a five-point scale. The average and standard deviation of each value are displayed on right side of each
score bar. There are no significantly statistical differences (all, _p_-values > 0.05). Inter-reader agreements of anatomical structures, perceived image noise, presence of imaging
artifacts on FSE and FSE-DNN were fair to moderate correlation (R2 = 0.73, 0.31, and 0.89, respectively; all, _p_ < 0.001). Inter-reader agreement on FSE and FSE-DNN showed good agreement
(R2 = 0.76; _p_ < 0.001). Representative images for comparable image quality are shown in Fig. 4. The accelerated image exhibited fewer motion-related artifacts, while the TSE-DNN image
displayed improved image quality. However, certain artifacts from parallel imaging persisted in the TSE-DNN image, as depicted in Fig. 5. DIAGNOSTIC PERFORMANCE FOR LIGAMENTOUS, MENISCAL,
AND CARTILAGINOUS LESIONS All FSE-conv and FSE-DNN images were rated of lesion detection by three interpreting musculoskeletal radiologists. In evaluation of lesion detection, the diagnostic
performances of FSE-DNN showed comparable results in ligamentous, meniscal, and cartilaginous lesions (Table 3, Fig. 6). Two of cartilage lesions was under-graded or over-graded (n = 2)
while there was no significant difference in other image sets (n = 43). Representative imaging examples for cartilage under-grading or over-grading are shown in Fig. 7. DISCUSSION In this
post-market multi-vendor study using commercially available DNN-based parallel imaging reconstruction, the FSE-DNN reconstruction of highly accelerated MRI scan reduced acquisition time by
an overall 41.0% for a 2D FSE image of the knee MRI. This algorithm is an image-based DNN-reconstruction, which does not require the _k_-space data nor MRI physics-related information such
as multi-channel coil geometry. An older model of this software has also shown promising generalizability results in pediatric brain32 and in prostate imaging as well33, indicating that an
image-based approach may enable accelerated MR exams in clinically used routine sequences with relatively simple modifications such as changing parallel imaging factors and the number of
phase encoding lines, etc. This software could restore highly accelerated images in a short time, displaying clinically acceptable image quality and comparable diagnostic performance. This
can be differentiated from physics-based (_k_-space-based) DNN reconstruction34 which is inherently closely associated with data acquisition methods and may require more computational power
and time. As a result, image-based algorithms can be relatively easily deployed in a variety of clinical settings with less MR vendor dependency, and they can even be applied retrospectively
to image data in the PACS server. However, further investigation is warranted to accurately compare the image qualities of image-based and k-space-based DNN reconstruction methods. The
strength of our study is that it is a prospective, multi-reader, multi-vendor study as a post-market surveillance. We applied the reconstruction algorithm to knee MRI sequences from all
three vendors, showing the possibility of application to multi-vendor MRI applications in radiology. With this strength of this FSE-DNN model, this image-based DNN reconstruction can be
easily employed in the radiologic workflow of multi-vendor MRI with various MRI parameters. By changing the parallel imaging factor or number of phase encoding steps of the conventional
routine MR sequence, which is easily applicable in the clinical MR imaging protocol, this DNN image reconstruction can reduce scan times with non-inferior image quality and comparable
diagnostic performance. In the quantitative evaluation, FSE-DNN reconstructed images showed higher SNR and CNR, corresponding with previous FSE-DNN studies35,36 and the same software32,33.
In subjective qualitative evaluation, FSE-DNN reconstructed images of accelerated FSE images showed non-inferiority against FSE-conventional images in terms of qualitative image quality
evaluation. Deep learning reconstruction can be employed in various accelerated imaging techniques37,38,39, such as parallel imaging, compressed sensing, or their combination. In our study,
we did not compare the combination of compressed sensing and parallel imaging with parallel imaging alone (e.g., CS-SENSE vs. SENSE or CS-SENSE vs. ASSET). Further study on deep learning
reconstruction comparison study on combination of CS and parallel imaging is needed in the future. In our study, overall image qualities of anatomical structures were better or equal in most
cases. By utilizing deep learning reconstruction of MRI, it is possible to reduce scan time and minimize patient movement, resulting in motion-less imaging. However, the artifacts related
with parallel imaging can be pronounced, and these artifacts may persist in some patients. This highlights the need for optimized MRI sequences tailored for accelerated MR imaging. Further
research in this direction is imperative in the future. This optimization of MRI sequences involves the choice of CS and the parallel imaging factor, and optimized _k-space_ trajectories. In
the diagnostic performance of lesion analysis, FSE-DNN reconstructed images showed non-inferiority compared to FSE-conventional images. In our study, no significant difference was observed
in the diagnostic performance between FSE-conventional and FSE-DNN images. In cartilage evaluation, FSE-DNN showed under-graded lesion (n = 1) and over-graded lesion (n = 1) in small numbers
(n = 2/45) among 45 image sets. However, there is no statistical difference between FSE-DNN reconstructed images and FSE-conventional images in this 45-case study. This under-grade or
over-grade of cartilage may have originated from acceleration artifacts and image degradation rather than the DNN- reconstruction in our early clinical validation with routine clinical MRI
protocols (Fig. 7). Cartilage under-grading on FSE-DNN could also have been affected by the amount of image denoising. This suggests the need for careful selection of acceleration method and
denoising settings for cartilage imaging, which may depend on imaging target structures. Conversely, cartilage could be over-graded from parallel imaging-related artifacts. In an
under-graded chondromalacia case, cartilage fissuring was smoothened on FSE-DNN images while cartilage signal was slightly enhanced in an over-graded chondromalacia case. This highlights the
necessity for MRI sequence optimization, particularly emphasizing the need for more precise learning when it comes to small structures like cartilage and structures influenced by MR signal
intensity. This acceleration optimization could be different depending on the target joint (e.g. a large off-center shoulder and a small extremity hand) and target structures such as
ligaments, bone marrow, meniscus, and cartilage. Further study involving a larger number of images is needed to validate this aspect. There were several limitations of this study as well.
First, the acquisition parameter modifications were not the same between the three vendors. For example, compressed sensing is routinely used in only one of the scanners, whereas the other
two scanners utilize parallel imaging only. We intended the DNN-based reconstruction application to the current MRI sequences, reflecting the clinical practice. Secondly, we set a denoising
level of low (51% reduction) according to a preceding internal study on noise reduction level for knee MRI. However, optimal image reduction level should be further investigated in clinical
MRI, which could depend on the scanner, imaging joint or target, and radio-frequency coil. Thirdly, our diagnostic evaluations were not confirmed arthroscopically in all patients. We
conducted this study with radiologists’ consensus as a gold standard of diagnostic performance. Despite evaluation in limited number of patients and limited pathologic findings, this
prospective study supports the possibility and the generalizability of DNN reconstruction of highly accelerated MRI to reduce the MRI scan time in clinical practice. Fourthly, the
measurements for SNR and CNR were conducted by conventional approach. While optimized methods are available for sensitivity map-based parallel imaging40,41, it was necessary to adhere to the
conventional approach for practical reasons, as multiple acquisitions were challenging to perform on actual patient images. And, in the context of deep learning-based MRI reconstruction, an
optimized method has not yet been established. Further research in this area is warranted in the future. In conclusion, DNN can be applied to accelerated knee imaging in multi-vendor MRI
scanners, with reduced scan time and comparable image quality. Cartilage grading could be under- or over-graded while good agreements in ligamentous and meniscal evaluations. Therefore, the
readers should be cautious in utilizing in DNN-accelerated MRI for some lesion evaluation. This study suggests the potential for routine MRI protocols applications to DNN-accelerated knee
MRI in clinical practice. DATA AVAILABILITY The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
ABBREVIATIONS * Acc: Accuracy * ACL: Anterior cruciate ligament * AUC: Area under curve * Ax T1W: Axial T1-weighted images * Ax T2W FS: Axial T2-weighted fat-saturated images * BM: Bone
Marrow * CNN: Convolutional neural network * Cor T2W FS: Coronal T2-weighted fat-saturated images * CS: Compressed sensing * DNN: Deep neural network * FCL: Fibular collateral ligament *
FSE: Fast spin-echo * HIPAA: Health insurance portability and accountability act * LM: Lateral meniscus * MCL: Medial collateral ligament * MM: Medial meniscus * MRI: Magnetic resonance
imaging * PACS: Picture archiving and communication system * PCL: Posterior cruciate ligament * PI: Parallel imaging * ROI: Region of interest * Sag T2W: Sagittal T2-weighted images * SD:
Standard deviation * Sen: Sensitivity * SNR: Signal-to-noise ratio * Spe: Specificity * SSIM: Structural similarity index * TSE: Turbo spin-echo REFERENCES * Nguyen, U. S. _et al._
Increasing prevalence of knee pain and symptomatic knee osteoarthritis: Survey and cohort data. _Ann. Intern. Med._ 155, 725–732. https://doi.org/10.7326/0003-4819-155-11-201112060-00004
(2011). Article PubMed PubMed Central Google Scholar * Oei, E. H., Nikken, J. J., Verstijnen, A. C., Ginai, A. Z. & Myriam Hunink, M. G. MR imaging of the menisci and cruciate
ligaments: A systematic review. _Radiology_ 226, 837–848. https://doi.org/10.1148/radiol.2263011892 (2003). Article PubMed Google Scholar * Kassarjian, A. _et al._ Guideliner for MR
Imaging of Sports Injuries. _European Society of Skeletal Radiology Sports Sub-commitee_ (2016). * (ACR), A. C. O. R., _MRI Exam-Specific Parameters: MSK Module (Revised 6–22–2022)_ (2022).
Available at: https://accreditationsupport.acr.org/support/solutions/articles/11000061021-mri-exam-specific-parameters-msk-module-revised-6-22-2022-#RevisionHistory8/31/2023). * Boutin, R.
D., Eshed, I., Kassarjian, A. & Vemuri, N. V. The global reading room: Knee MRI protocols. _AJR Am. J. Roentgenol._ 219, 347–348. https://doi.org/10.2214/AJR.21.27238 (2022). Article
PubMed Google Scholar * Smith, C. _et al._ Diagnostic efficacy of 3-T MRI for knee injuries using arthroscopy as a reference standard: A meta-analysis. _AJR Am. J. Roentgenol._ 207,
369–377. https://doi.org/10.2214/AJR.15.15795 (2016). Article PubMed Google Scholar * Iuga, A. I. _et al._ Accelerated MRI of the knee. Quality and efficiency of compressed sensing. _Eur
J Radiol_ 132, 109273. https://doi.org/10.1016/j.ejrad.2020.109273 (2020). Article PubMed Google Scholar * Chaudhari, A. S. _et al._ Combined 5-minute double-echo in steady-state with
separated echoes and 2-minute proton-density-weighted 2D FSE sequence for comprehensive whole-joint knee MRI assessment. _J. Magn. Reson. Imaging_ 49, e183–e194.
https://doi.org/10.1002/jmri.26582 (2019). Article PubMed Google Scholar * Garwood, E. R., Recht, M. P. & White, L. M. Advanced imaging techniques in the knee: Benefits and
limitations of new rapid acquisition strategies for routine knee MRI. _AJR Am. J. Roentgenol._ 209, 552–560. https://doi.org/10.2214/AJR.17.18228 (2017). Article PubMed Google Scholar *
Alaia, E. F. _et al._ Comparison of a fast 5-min knee MRI protocol with a standard knee MRI protocol: A multi-institutional multi-reader study. _Skeletal. Radiol._ 47, 107–116.
https://doi.org/10.1007/s00256-017-2780-9 (2018). Article PubMed Google Scholar * Pruessmann, K. P., Weiger, M., Scheidegger, M. B. & Boesiger, P. SENSE: Sensitivity encoding for fast
MRI. _Magn. Reson. Med._ 42, 952–962 (1999). Article CAS PubMed Google Scholar * Deshmane, A., Gulani, V., Griswold, M. A. & Seiberlich, N. Parallel MR imaging. _J. Magn. Reson.
Imaging_ 36, 55–72. https://doi.org/10.1002/jmri.23639 (2012). Article PubMed PubMed Central Google Scholar * Vasanawala, S. S. _et al._ Improved pediatric MR imaging with compressed
sensing. _Radiology_ 256, 607–616. https://doi.org/10.1148/radiol.10091218 (2010). Article PubMed PubMed Central Google Scholar * Jaspan, O. N., Fleysher, R. & Lipton, M. L.
Compressed sensing MRI: A review of the clinical literature. _Br. J. Radiol._ 88, 20150487. https://doi.org/10.1259/bjr.20150487 (2015). Article PubMed PubMed Central Google Scholar *
Lustig, M., Donoho, D. & Pauly, J. M. Sparse MRI: The application of compressed sensing for rapid MR imaging. _Magn. Reson. Med._ 58, 1182–1195. https://doi.org/10.1002/mrm.21391 (2007).
Article PubMed Google Scholar * Knoll, F. _et al._ fastMRI: A publicly available raw k-Space and DICOM dataset of knee images for accelerated MR image reconstruction using machine
learning. _Radiol. Artif. Intell._ 2, e190007. https://doi.org/10.1148/ryai.2020190007 (2020). Article PubMed PubMed Central Google Scholar * Eo, T. _et al._ KIKI-net: Cross-domain
convolutional neural networks for reconstructing undersampled magnetic resonance images. _Magn. Reson. Med._ 80, 2188–2201. https://doi.org/10.1002/mrm.27201 (2018). Article CAS PubMed
Google Scholar * Akai, H. _et al._ Commercially available deep-learning-reconstruction of MR imaging of the knee at 1.5T has higher image quality than conventionally-reconstructed imaging
at 3T: A normal volunteer study. _Magn. Reson. Med. Sci._ https://doi.org/10.2463/mrms.mp.2022-0020 (2022). Article PubMed PubMed Central Google Scholar * Artificial Intelligence/Machine
Learning (AI/ML)-Based Software as a Medical Device (SaMD) Action Plan. U.S. Food & Drug Administration Center for Devices and Radiological Health.
https://www.fda.gov/media/145022/download. * Bash, S. _et al._ Deep learning enables 60% accelerated volumetric brain MRI while preserving quantitative performance: A prospective,
multicentre, multireader trial. _AJNR Am. J. Neuroradiol._ 42, 2130–2137. https://doi.org/10.3174/ajnr.A7358 (2021). Article CAS PubMed PubMed Central Google Scholar * Pasumarthi, S.
_et al._ A generic deep learning model for reduced gadolinium dose in contrast-enhanced brain MRI. _Magn. Reson. Med._ 86, 1687–1700. https://doi.org/10.1002/mrm.28808 (2021). Article CAS
PubMed Google Scholar * Bash, S. _et al._ Deep learning image processing enables 40% faster spinal MR scans which match or exceed quality of standard of care : A prospective multicenter
multireader study. _Clin. Neuroradiol._ 32, 197–203. https://doi.org/10.1007/s00062-021-01121-2 (2022). Article CAS PubMed Google Scholar * Ronneberger, O., Fischer, P. & Brox, T.
U-Net: Convolutional networks for biomedical image segmentation. _Lect. Notes Comput. Sci._ 9351, 234–241. https://doi.org/10.1007/978-3-319-24574-4_28 (2015). Article Google Scholar *
Kingma, D. P., Ba, J. Adam: A method for stochastic optimization, in _Proceedings of the International Conference on Learning Representations_ 1–15 (2015). * Crues, J. V. 3rd., Mink, J.,
Levy, T. L., Lotysch, M. & Stoller, D. W. Meniscal tears of the knee: Accuracy of MR imaging. _Radiology_ 164, 445–448. https://doi.org/10.1148/radiology.164.2.3602385 (1987). Article
PubMed Google Scholar * De Smet, A. A. _et al._ MR diagnosis of meniscal tears of the knee: Importance of high signal in the meniscus that extends to the surface. _AJR Am. J. Roentgenol._
161, 101–107. https://doi.org/10.2214/ajr.161.1.8517286 (1993). Article PubMed Google Scholar * Chen, W. T., Shih, T. T., Tu, H. Y., Chen, R. C. & Shau, W. Y. Partial and complete
tear of the anterior cruciate ligament. _Acta Radiol._ 43, 511–516 (2002). PubMed Google Scholar * Sonin, A. H., Fitzgerald, S. W., Hoff, F. L., Friedman, H. & Bresler, M. E. MR
imaging of the posterior cruciate ligament: Normal, abnormal, and associated injury patterns. _Radiographics_ 15, 551–561. https://doi.org/10.1148/radiographics.15.3.7624562 (1995). Article
CAS PubMed Google Scholar * Ruiz, M. E. & Erickson, S. J. Medial and lateral supporting structures of the knee. Normal MR imaging anatomy and pathologic findings. _Magn. Reson.
Imaging Clin. N. Am._ 2, 381–399 (1994). Article CAS PubMed Google Scholar * Potter, H. G., Linklater, J. M., Allen, A. A., Hannafin, J. A. & Haas, S. B. Magnetic resonance imaging
of articular cartilage in the knee. An evaluation with use of fast-spin-echo imaging. _J. Bone Joint Surg. Am._ 80, 1276–1284 (1998). Article CAS PubMed Google Scholar * Suh, J. S., Lee,
S. H., Jeong, E. K. & Kim, D. J. Magnetic resonance imaging of articular cartilage. _Eur. Radiol._ 11, 2015–2025. https://doi.org/10.1007/s003300100911 (2001). Article CAS PubMed
Google Scholar * Jung, W., Kim, J., Ko, J., Jeong, G. & Kim, H. G. Highly accelerated 3D MPRAGE using deep neural network-based reconstruction for brain imaging in children and young
adults. _Eur. Radiol._ 32, 5468–5479. https://doi.org/10.1007/s00330-022-08687-6 (2022). Article CAS PubMed Google Scholar * Jung, W., Kim, E. H., Ko, J., Jeong, G. & Choi, M. H.
Convolutional neural network-based reconstruction for acceleration of prostate T2 weighted MR imaging: A retro- and prospective study. _Br. J. Radiol._ 95, 20211378.
https://doi.org/10.1259/bjr.20211378 (2022). Article PubMed Google Scholar * Lebel, R. M. Performance characterization of a novel deep learning-based MR image reconstruction pipeline.
arXiv preprint arXiv:2008.06559 (2020). * Ueda, T. _et al._ Deep learning reconstruction of diffusion-weighted MRI improves image quality for prostatic imaging. _Radiology_ 303, 373–381.
https://doi.org/10.1148/radiol.204097 (2022). Article PubMed Google Scholar * Obama, Y. _et al._ MR imaging for shoulder diseases: Effect of compressed sensing and deep learning
reconstruction on examination time and imaging quality compared with that of parallel imaging. _Magn. Reson. Imaging_ 94, 56–63. https://doi.org/10.1016/j.mri.2022.08.004 (2022). Article
PubMed Google Scholar * Qiu, D., Zhang, S., Liu, Y., Zhu, J. & Zheng, L. Super-resolution reconstruction of knee magnetic resonance imaging based on deep learning. _Comput. Methods
Prog. Biomed._ 187, 105059. https://doi.org/10.1016/j.cmpb.2019.105059 (2020). Article Google Scholar * Hammernik, K. _et al._ Learning a variational network for reconstruction of
accelerated MRI data. _Magn. Reson. Med._ 79, 3055–3071. https://doi.org/10.1002/mrm.26977 (2018). Article PubMed Google Scholar * Johnson, P. M. _et al._ Deep learning reconstruction
enables prospectively accelerated clinical knee MRI. _Radiology_ 307, e220425. https://doi.org/10.1148/radiol.220425 (2023). Article PubMed Google Scholar * Goerner, F. L. & Clarke,
G. D. Measuring signal-to-noise ratio in partially parallel imaging MRI. _Med. Phys._ 38, 5049–5057. https://doi.org/10.1118/1.3618730 (2011). Article PubMed PubMed Central Google Scholar
* Yu, J., Agarwal, H., Stuber, M. & Schar, M. Practical signal-to-noise ratio quantification for sensitivity encoding: Application to coronary MR angiography. _J. Magn. Reson. Imaging_
33, 1330–1340. https://doi.org/10.1002/jmri.22571 (2011). Article PubMed PubMed Central Google Scholar Download references ACKNOWLEDGEMENTS AIRS Medical provided financial support for
this prospective study. The authors had control of the data and the information submitted for publication. We would like to thank Jinseob Kim (Zarathu Co., Ltd, Seoul, Korea) for statistical
advice. This study was supported by a faculty research grant of Yonsei University College of Medicine (6-2017-0071). AUTHOR INFORMATION AUTHORS AND AFFILIATIONS * Department of Radiology,
Research Institute of Radiological Science, and Center for Clinical Imaging Data Science (CCIDS), Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul, 03722, Korea
Joohee Lee, Jiwoo Park, Sungjun Kim, Yunjin Im, Nim Lee, Ho-Taek Song & Young Han Lee * Department of Orthopaedic Surgery, Yonsei University College of Medicine, Seoul, Korea Min Jung
Authors * Joohee Lee View author publications You can also search for this author inPubMed Google Scholar * Min Jung View author publications You can also search for this author inPubMed
Google Scholar * Jiwoo Park View author publications You can also search for this author inPubMed Google Scholar * Sungjun Kim View author publications You can also search for this author
inPubMed Google Scholar * Yunjin Im View author publications You can also search for this author inPubMed Google Scholar * Nim Lee View author publications You can also search for this
author inPubMed Google Scholar * Ho-Taek Song View author publications You can also search for this author inPubMed Google Scholar * Young Han Lee View author publications You can also
search for this author inPubMed Google Scholar CONTRIBUTIONS J.L., J.P., and Y.H.L. wrote the main manuscript. M.J. and Y.H.L. completed the data collection. J.L., J.P., N.L., and Y.L.
performed the imaging analysis. M.J., S.K., and H.-T.S. edited the manuscripts. All authors reviewed the final manuscript. CORRESPONDING AUTHOR Correspondence to Young Han Lee. ETHICS
DECLARATIONS COMPETING INTERESTS The authors declare no competing interests. ADDITIONAL INFORMATION PUBLISHER'S NOTE Springer Nature remains neutral with regard to jurisdictional claims
in published maps and institutional affiliations. RIGHTS AND PERMISSIONS OPEN ACCESS This article is licensed under a Creative Commons Attribution 4.0 International License, which permits
use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the
Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless
indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory
regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit
http://creativecommons.org/licenses/by/4.0/. Reprints and permissions ABOUT THIS ARTICLE CITE THIS ARTICLE Lee, J., Jung, M., Park, J. _et al._ Highly accelerated knee magnetic resonance
imaging using deep neural network (DNN)–based reconstruction: prospective, multi-reader, multi-vendor study. _Sci Rep_ 13, 17264 (2023). https://doi.org/10.1038/s41598-023-44248-7 Download
citation * Received: 12 March 2023 * Accepted: 05 October 2023 * Published: 12 October 2023 * DOI: https://doi.org/10.1038/s41598-023-44248-7 SHARE THIS ARTICLE Anyone you share the
following link with will be able to read this content: Get shareable link Sorry, a shareable link is not currently available for this article. Copy to clipboard Provided by the Springer
Nature SharedIt content-sharing initiative