Play all audios:
ABSTRACT In the past decades, image-guided surgery has evolved rapidly. In procedures with a relatively fixed target area, like neurosurgery and orthopedics, this has led to improved patient
outcomes. In cancer surgery, intraoperative guidance could be of great benefit to secure radical resection margins since residual disease is associated with local recurrence and poor
survival. However, most tumor lesions are mobile with a constantly changing position. Here, we present an innovative technique for real-time tumor tracking in cancer surgery. In this study,
we evaluated the feasibility of real-time tumor tracking during rectal cancer surgery. The application of real-time tumor tracking using an intraoperative navigation system is feasible and
safe with a high median target registration accuracy of 3 mm. This technique allows oncological surgeons to obtain real-time accurate information on tumor location, as well as critical
anatomical information. This study demonstrates that real-time tumor tracking is feasible and could potentially decrease positive resection margins and improve patient outcome. SIMILAR
CONTENT BEING VIEWED BY OTHERS ENABLING THE FUTURE OF COLONOSCOPY WITH INTELLIGENT AND AUTONOMOUS MAGNETIC MANIPULATION Article 12 October 2020 MAGNETIC SURGICAL MARKER NAVIGATION FOR
EXCISION OF NON-PALPABLE ULTRASOUND VISIBLE BREAST LESIONS: FIRST 200 CASES IN A FRENCH CANCER CENTER Article Open access 10 February 2025 MOTOR-FREE TELEROBOTIC ENDOMICROSCOPY FOR STEERABLE
AND PROGRAMMABLE IMAGING IN COMPLEX CURVED AND LOCALIZED AREAS Article Open access 03 September 2024 INTRODUCTION In the past two decades, image-guided surgery has evolved rapidly. The
technique enables real-time visualization of surgical instrumentation with respect to intraoperative anatomy based on preoperative imaging. A prerequisite of the current surgical navigation
systems is a fixed target area since registration of the preoperative images with the intraoperative anatomy is based on this rigid, fixed position. In view of this condition, the technique
was initially developed in neurosurgery, in which the soft-tissue component has a fixed position with respect to the bone. The main goal was to improve precise targeting of the tumor margins
and improve the safety of the surgical procedures. Image-guided neurosurgery evolved from frame-based stereotaxic to frameless navigation1. Nowadays, multiple studies showed that
image-guided neurosurgery is feasible and improves the extent of resection of malignant brain tumors2,3. Additionally, the technique expanded to orthopedic and head and neck surgery where it
showed to be of added value with improved accurate placement of surgical implants4,5. Besides these applications, intraoperative guidance could be of great benefit to many other surgical
oncology procedures not confined to a rigid surrounding. During such procedures, surgeons rely on palpation and visual inspection to distinguish tumorous from healthy tissue. This can be
challenging and may result in incomplete removal of the tumor tissue, associated with local recurrence and poor overall survival6,7,8,9. To make navigation useful under such circumstances
and improve surgical outcomes, the technique has to account for the changing position of the tumor by real-time tumor tracking during surgery. In this study, we present a new technique that
brings surgical navigation to the next stage of development and make it suitable for a wide range of applications in surgical oncology. To this end, we developed an innovative technology
that allows for real-time tumor tracking during surgery. Here, we report on the feasibility and accuracy of this innovative technique in patients undergoing rectal surgery. In rectal cancer
surgery, total mesorectal excision (TME) is the golden standard. TME can be challenging with limited visualization and surgical access due to the anatomical location of the tumor. Because of
this, the amount of tumor positive resection margins still remains 10–15%10,11. Therefore, better intraoperative guidance would be of great benefit to secure radical resection margins. The
study was performed in two phases. In the first phase (feasibility phase), feasibility and safety of the study workflow were investigated, the possibility of real-time tumor tracking with
the navigation system was assessed and possible improvements of the clinical workflow were identified. In the second phase (test phase), the ultimate accuracy of the real-time tracking was
measured. RESULTS FEASIBILITY PHASE Between November 2016 and March 2018, 15 patients with histological proven rectal cancer were included (Table 1). In all patients, a patient-specific
three-dimensional (3D) model was created based on preoperative imaging. The time to segment the preoperative images took around 2 h. A computer screen in the OR showed the 3D model, three
orthogonal views (axial, sagittal, and coronal) of the preoperative images, and the tracked tumor and pointer position during surgery (Fig. 1). A schematic overview of the position of the
tumor sensor that adjusted for real-time tumor movements is shown in Fig. 2 and an example of the tracked tumor movement in Fig. 3. In the feasibility phase, 11 patients (73.3%) successfully
completed the entire investigational workflow and received rectal surgery with the navigation technique. In four patients (26.7%) the tumor sensor was damaged during placement and
discontinued working. Therefore, the sensor was inserted in a silicone round surgical wound drain (3.3 mm) for protection before placement into the rectum in the subsequent patients
(starting at patient 9). During the feasibility phase, tumor registration was challenging due to the low soft-tissue contrast on the CBCT. Visual evaluation of the registration during
surgery, when the patient was in Trendelenburg position, showed that manual correction of the system was necessary in four patients with an average of 3.8 mm in the left–right direction, 9.4
mm caudal–cranial direction, and 5.5 mm in anterior–posterior direction. In one patient, the intratumoral fiducial was lost after placement. Registration was based on the remaining two
fiducials in this patient. No navigation-related complications were observed. Treatment-related characteristics are shown in Table 2. Placement of the patient trackers, positioning the tumor
sensor, and obtaining the intraoperative CBCT scan added on average 31 min (range: 21–46) to the total surgical procedure time. TEST PHASE Since tumor registration in the feasibility phase
was challenging, patients received in the test phase fiducial markers close to the rectal tumor before the planning CT scan, at least 1 week before surgery. In the OR, registration of the
tumor was performed based on the fiducial markers which allowed for a more accurate intraoperative tumor registration and tracking. During the test phase, the definite protocol was evaluated
in 16 patients (Table 1). Of these patients, 14 patients received rectal surgery with the navigation system. In two patients, navigation was not used during surgery. In the first patient,
rectal surgery was canceled due to unexpected progression of metastatic disease. The second patient suffered from an intra-abdominal infection possibly related to the fiducial placement and
underwent early resection without navigation. Treatment-related characteristics are shown in Table 2. During surgery with the patient in the Trendelenburg position, adjustment of the
registration was performed in six patients. The average correction was 2.3 mm in the left–right direction, 13.0 mm caudal–cranial direction, and 3.8 mm in anterior–posterior direction. In
one patient, the intratumoral fiducial was lost after placement, resulting in a tumor registration based on the remaining two fiducials in this patient. In the test phase, placement of the
patient trackers, positioning of the tumor sensor, and obtaining the intraoperative CBCT scan added on average 21 min (range: 15–30) to the total surgical procedure time. Real-time tracking
of the rectal tumor was feasible in 13 of the 14 patients (92.9%). In one patient, despite manual correction during the visual evaluation, no accurate registration could be achieved and the
registration error remained >2 cm. Therefore, no interpretable navigation data could be collected. No navigation-related complications were observed during surgery. Accuracy of the
navigation setup was determined by correlation of the position of the proximal tumor margin, as determined during navigation and marked by surgical clips, with the position of the proximal
tumor margin as determined on the pathology specimen. Accuracy measurements could be achieved in 11 patients (84.6%) of the 13 patients with navigation data receiving rectal surgery. In one
patient pathological correlation of tumor margins was not possible due to a complete pathological response (T0N0) after neoadjuvant treatment and in one other patient because the surgical
clips to mark the proximal tumor margin during navigation fell off the rectal resection specimen during surgical removal. The median absolute distance between the surgical clips and the
proximal tumor border at histopathology was 3 mm (IQR: 2–10 mm). DISCUSSION Image-guided surgical navigation has become widespread in head and neck surgery, neurosurgery, and orthopedics. In
these fields, the technique is feasible due to the most rigid, bone-defined target area. To make surgical navigation a useful technique for cancer surgery in general, where tumor lesions
often shift during surgery, real-time tumor tracking is necessary. In this article, we present a navigation system that fulfills this need and performs real-time tumor tracking during the
course of surgery. The system was tested in rectal cancer surgery in both open and laparoscopic setting. This prospective study shows that real-time tumor tracking with surgical navigation
is feasible, safe, and accurate. During the feasibility phase, improvements in the workflow were made to enhance the practicality, speed, and accuracy of the tumor registration. After
optimization of the workflow, a high median accuracy of 3 mm was obtained for surgical navigation in the test phase of the study. No navigation-related complications were observed during
surgery and the total extra time decreased with one-third towards the end of the study to 21 min. The accuracy obtained is clinically relevant and warrants further clinical implementation.
In the past years, some limited case series evaluated the feasibility of surgical navigation in pelvic surgery. The group of Atallah et al.12,13,14 published three articles, including five
patients in total, on stereotactic navigation using an optical tracking system for transanal minimal invasive surgery—TME (TAMIS-TME). In all three studies, they concluded that stereotactic
navigation for TAMIS-TME was feasible and safe with an accuracy of around 4 mm. However, the accuracy of the system did not assume tumor or patient movement during the procedure as the
trackers were mounted on to the bedrail. Kwak et al.15 also performed navigation in pelvic surgery in two patients, stating that surgical navigation was feasible with acceptable accuracy of
the target. Although both groups showed encouraging results, no real-time tracking of the rectal tumor was performed and the rectal tumor was assumed to be fixed in the same position during
the procedure. The main challenge of the current study was to track tumor movement in real time in every possible direction. This could be accomplished by introducing a six degree of freedom
sensor close to the tumor lesion. The current results confirm earlier results from an ex vivo rectum study by Wagner et al.16. They developed a realistic rectum phantom to test motion and
deformation using a tracking sensor. Multiple models were created to translate sensor movement to rectum wall deformations. In the most complex model, 33 parameters were taken into account
which resulted in a target registration error of 2.9 ± 1.4 mm. Our results are in accordance with the results obtained in these phantom experiments. A limiting factor for the accuracy
measurements in the current in vivo application for rectal cancer may be that all patients in the current study were operated after neoadjuvant radiotherapy with a long interval to surgery.
This makes interpretation of the preoperative imaging more challenging. Magnetic resonance imaging (MRI) is the most advanced staging technique in rectal cancer and routinely used for
preoperative staging and restaging17,18. Unfortunately, interpretation of MRI is known to be less reliable when patients are pretreated with chemoradiation or radiotherapy. Multiple studies
evaluated MRI for restaging and showed overall a low sensitivity, which is mainly caused by difficulties to differentiate areas of vital residual tumor within radiation-induced
fibrosis19,20. In this study, all segmentations of the tumor and other critical structures were validated by an experienced radiologist or a rectal cancer MRI expert and the operating
surgeon. Despite this challenging task, a median navigation accuracy of 3 mm was established. Besides the interpretation of the MRI, some other factors may have influenced the accuracy. For
example, the obtained accuracy may be dependent on the positioning and registration of the fiducials. Fiducial placement was performed using a sigmoidoscopy in combination with endoscopic
ultrasonography (EUS) but in smaller limited-volume tumors, adequate positioning of the fiducials on the proximal and distal tumor was challenging. Furthermore, it remains uncertain if the
position of the fiducials was indeed rigid with respect to the tumor. Another factor that might have influenced the accuracy is the pathological processing of the rectal specimen. The
distance between the surgical clips and the proximal tumor border was measured manually by the pathologist and reproduction of the axial tumor plane was in some patients complex. Manual
correction of the registration was performed in 10 out of 29 patients. When corrections were applied, these were almost exclusively related to the caudal–cranial axis (9 out of 10). A
possible explanation for the inaccuracy in this direction may be found in the way patients were tracked during surgery. Patient trackers were taped to the skin close to the bone to correct
for patient movement with respect to the table and field generator. However, some movement of the skin and the patient trackers with respect to the pelvis may be inevitable, especially when
the patient is moved to the Trendelenburg position during surgery. Obtaining the CBCT scan after the patient is positioned to Trendelenburg is unfortunately not possible since a security
lock in the CBCT scanner prevents scanning when the bed is not positioned straight. However, this slight movement could easily be corrected for by visual correction of the registration after
the patient was positioned in Trendelenburg. During the study, adjustments were performed improving the practicality and the accuracy of the tumor registration. However, positioning and
securing the tumor sensor against the rectal tumor remains inconvenient and adds additional time to the anesthesia. A solution may be found in the use of wireless EM-tracked
transponders21,22,23. These wireless transponders could be implanted near the tumor the day before surgery when the patient would be admitted to the hospital, and their position could be
visualized by CT the same day. During the OR no additional imaging would be necessary, and any shift of tumor location could simply be tracked in real time by tracking the responder
position. In conclusion, this study shows that real-time tumor tracking for cancer surgery is feasible and safe. The accuracy of the system is high, which justifies broader clinical
implementation. Further studies will be initiated to improve the workflow and prove clinical benefit. METHODS CLINICAL TRIAL DESIGN AND PATIENT INCLUSION This prospective feasibility study
was conducted in The Netherlands Cancer Institute from November 2016 to August 2019. The study was approved by the institutional review board in August 2016 (NL57251.031.16/N16TRS). The
study was conducted according to the principles of the Declaration of Helsinki and in accordance with the Medical Research Involving Human Subjects Act (WMO). Inclusion criteria were
patients with rectal cancer scheduled for open or laparoscopic low anterior resection (LAR) or abdominoperineal resection (APR), patient age ≥18 years, tumor distance from anal verge <10
cm (based on preoperative imaging). Exclusion criteria were contraindications for intravenous contrast administration (allergy or severely impaired kidney function), cardiac pacemaker and
metal implants in the pelvic area causing artifacts on the preoperative imaging. All participating patients gave written informed consent. The study was registered at www.trialregister.nl
(identifier: NL7666). The study was performed in two phases. In the first phase (feasibility phase), 15 patients were included. The feasibility and safety of the study workflow were
investigated, the possibility of real-time tumor tracking with the navigation system was assessed, and possible improvements of the study protocol were identified. After optimization of the
study protocol, the second phase (test phase) was initiated. In this test phase, 16 patients were included and the accuracy of the real-time tracking was determined. NAVIGATION SYSTEM An NDI
Aurora V2 electromagnetic (EM) tracking system (Northern Digital Inc., Waterloo, Ontario, Canada) was used to link preoperative imaging data to the intraoperative patient setup. Patients
were positioned on an operating table including a tabletop field generator (TTFG) with an oval EM field of 42 × 60 × 60 cm. Patient trackers (Philips Traxtal/percunav, Philips, Best, The
Netherlands) with EM sensors were used to determine the position of the patient during surgery and a tracking sensor was placed against the tumor to adjust for real-time tumor movements.
Within a hybrid operating room, a cone-beam computed tomography (CBCT) scan (Philips Allura FD20 XperCT; Philips) was performed after sensor placement but before the start of surgery.
Acquired images were registered to the preoperative diagnostic planning CT scan using the in-house developed navigation software. During surgery, an EM-tracked surgical pointer (NDI) was
available to validate the accuracy of the registration. WORKFLOW IN FEASIBILITY PHASE A preoperative diagnostic multiphase contrast-enhanced abdominal CT scan (planning CT with early
arterial and excretion phase)—acquired for the purpose of this study one day before surgery—and diagnostic MRI scans were used to create a patient-specific 3D model. The 3D model consisting
of bones, arteries, veins, ureters, rectum, and tumor was created using semi-automated in-house developed segmentation software. Tumor delineation was supervised by an experienced
radiologist or study investigator with extensive experience in rectal cancer MRI. Resulting 3D models were evaluated by the operating surgeon before surgery. In the operation room (OR),
three patients trackers (PercuNav Patient Tracker, Philips) were taped to the skin of the patient. One patient tracker was positioned at the level of the anterior superior iliac spine and
the remaining two left and right of the spine in the lumbar curvature. Under anesthesia, a rectal examination was performed by the surgeon. With a proctoscope, a wired 6 degree of freedom
EM-sensor (Flextube, Northern Digital Inc., Waterloo, Ontario, Canada) was fixated on the tumor using surgical glue. The patient was placed in surgical position and a CBCT image of the
pelvic area including the patient trackers was acquired. Because of the high mobility of the rectum, accurate registration of the intraoperative CBCT scan to the planning CT consisted of two
steps: (I) rigid registration of the (semi) rigid parts of the model (i.e., bones and main vessels) and (II) deformable registration of the tumor. First, the bony anatomy on the
intraoperative CBCT scan was linked to the 3D model by rigid registration. Subsequently, maintaining the registration of the bones and semi-rigid anatomy unchanged, a separate deformable
tumor registration was performed based on gray values. The resulting navigation information consisted of the rigid anatomy of the pelvis plus a non-rigid representation of the tumor. The
tumor sensor and pointer were automatically real-time tracked by the system and visible on a computer screen in the OR24. On this screen, the 3D model and three orthogonal views (axial,
sagittal, and coronal) of the obtained images were shown (Fig. 1). An example of the tracked tumor movement during surgery is shown in Fig. 3. At the start of surgery, the actual position of
the pointer was visually compared to the pointer position in the navigation system. In case of a visual inaccuracy of ≥2 mm, the registration was adjusted accordingly. During further
surgery, visual evaluation of the registration accuracy was performed by placing the tip of the pointer at anatomic landmarks like the sacral promontory, ureters, or the common iliac artery
bifurcation. If necessary, manual correction was performed on both the rigid anatomy as well as the tumor. WORKFLOW IN TEST PHASE For the test phase, the clinical workflow was changed with
regard to tumor sensor placement and registration of the tumor. To secure protection for breakage and guarantee a more stable tumor sensor position, the sensor was inserted in a round
silicone surgical wound drain (3.3 mm) before placement into the rectum. After positioning the tip of the sensor against the rectal tumor, the silicone wound drain was fixated to the skin in
the perineal area with stwo sutures. In the feasibility phase, tumor registration was challenging due to the low soft-tissue contrast on the CBCT. In the test phase, to improve tumor
registration, patients received fiducial markers close to the rectal tumor before the diagnostic CT scan, at least 1 week before surgery. These markers were used to facilitate tumor
registration between the low soft-tissue contrast CBCT images and the high soft-tissue contrast of the planning CT. Fiducial placement in the rectum is often used for tumor localization
during radiotherapy and has been reported as safe with excellent visibility on CT25,26,27. Patients received an enema before the procedure. Using flexible sigmoidoscopy in combination with
EUS, an experienced gastroenterologist inserted three fiducials through the bowel wall in the mesorectal fat (one on the proximal and distal tumor border and one halfway the tumor). The
locations of the fiducials were assumed to be rigid with respect to the tumor. Two types of fiducial markers were used: Gold Anchor GA150-20 0.28 mm × 20 mm unfolded length (Naslund Medical
AB, Huddinge, Sweden) and Visicoil 0.5 mm × 5 mm (Core Oncology, Santa Barbara, California). The Gold Anchors were folded when inserted. Preoperative fiducial placement by the
gastroenterologist took 20 min on average. In the OR, after the bone to bone registration, registration of the tumor was performed based on the fiducial markers. This allowed for a more
accurate intraoperative tumor registration and tracking. ACCURACY MEASUREMENTS In the test phase, an axial plane was added to the 3D model to accentuate the proximal tumor border. During
surgery, the surgeon used the navigation interface to place two titanium surgical clips on the surface of the rectum in the plane of the proximal tumor border as given by the 3D model. After
the specimen was resected, it was taken to the pathology department and processed according to standard protocol28. The distance between the surgical clips and the proximal tumor border was
macroscopically reported and verified with histopathology. If the surgeon was not able to reach the proximal tumor border (as indicated by the navigation system) with the surgical
instruments due to narrow pelvic space, surgical clips were placed as close as possible cranial of the axial plane. The difference between the distance of the clips relative to the proximal
tumor border measured by the navigation system and the pathology measurements was used as the accuracy value. OUTCOME PARAMETERS The main goal of the feasibility phase was to assess if
real-time tracking of the rectal tumor using the in-house developed EM navigation system was feasible and safe. Feasibility was defined as successful completion of the whole investigational
workflow resulting in continuous delivery of interpretable navigation data for rectal surgery. In addition, the amount of extra time needed to use the system during surgery was evaluated.
The main goal of the test phase was to evaluate the accuracy of the system, which was validated by the distance (in mm) between the intraoperatively placed surgical clips, illustrating the
proximal tumor border according to the 3D model, and the proximal tumor border as determined by pathology. In this phase, the amount of extra time was also reported. All analyses were
performed in SPSS version 24.0® (IBM Corporation, Armonk, NY, USA). DATA AVAILABILITY The data used for the current study are available from the corresponding author (E.N.D.K.) upon
reasonable request and through collaborative investigations. CODE AVAILABILITY The software used for the current study are available from the corresponding author (E.N.D.K.) upon reasonable
request and through collaborative investigations. REFERENCES * Schulz, C., Waldeck, S. & Mauer, U. M. Intraoperative image guidance in neurosurgery: development, current indications, and
future trends. _Radiol. Res. Pract._ 2012, 197364–197364 (2012). PubMed PubMed Central Google Scholar * Claus, E. B. et al. Survival rates in patients with low-grade glioma after
intraoperative magnetic resonance image guidance. _Cancer_ 103, 1227–1233 (2005). Article Google Scholar * Senft, C. et al. Intraoperative MRI guidance and extent of resection in glioma
surgery: a randomised, controlled trial. _Lancet Oncol._ 12, 997–1003 (2011). Article Google Scholar * Aschendorff, A. et al. Radiologically assisted navigation in cochlear implantation
for X-linked deafness malformation. _Cochlear Implants Int._ 10, 14–18 (2009). Article Google Scholar * Tormenti, M. J. et al. Intraoperative computed tomography image-guided navigation
for posterior thoracolumbar spinal instrumentation in spinal deformity surgery. _Neurosurgical Focus_ 28, E11 (2010). Article Google Scholar * Adam, I. J. et al. Role of circumferential
margin involvement in the local recurrence of rectal cancer. _Lancet_ 344, 707–711 (1994). Article CAS Google Scholar * El-Sherif, A. et al. Margin and local recurrence after sublobar
resection of non-small cell lung cancer. _Ann. Surg. Oncol._ 14, 2400–2405 (2007). Article Google Scholar * Houssami, N., Macaskill, P., Marinovich, M. L. & Morrow, M. The association
of surgical margins and local recurrence in women with early-stage invasive breast cancer treated with breast-conserving therapy: a meta-analysis. _Ann. Surg. Oncol._ 21, 717–730 (2014).
Article Google Scholar * Shindoh, J. et al. Complete removal of the tumor-bearing portal territory decreases local tumor recurrence and improves disease-specific survival of patients with
hepatocellular carcinoma. _J. Hepatol._ 64, 594–600 (2016). Article Google Scholar * Bonjer, H. J. et al. A randomized trial of laparoscopic versus open surgery for rectal cancer. _N.
Engl. J. Med._ 372, 1324–1332 (2015). Article CAS Google Scholar * Rickles, A. S. et al. High rate of positive circumferential resection margins following rectal cancer surgery: a call to
action. _Ann. Surg._ 262, 891–898 (2015). Article Google Scholar * Atallah, S., Martin-Perez, B. & Larach, S. J. T. i.c. Image-guided real-time navigation for transanal total
mesorectal excision: a pilot study. _Tech. Coloproctol._ 19, 679–684 (2015). Article CAS Google Scholar * Atallah, S., Nassif, G. & Larach, S. J. S. E. Stereotactic navigation for
TAMIS-TME: opening the gateway to frameless, image-guided abdominal and pelvic surgery. _Surg. Endosc._ 29, 207–211 (2015). Article Google Scholar * Atallah, S. et al. Robotic-assisted
stereotactic real-time navigation: initial clinical experience and feasibility for rectal cancer surgery. _Tech. Coloproctol._ 23, 53–63 (2019). Article CAS Google Scholar * Kwak, J. M.
et al. Stereotactic pelvic navigation with augmented reality for transanal total mesorectal excision. _Dis. Colon Rectum_ 62, 123–129 (2019). Article Google Scholar * Wagner, M. et al.
Electromagnetic organ tracking allows for real-time compensation of tissue shift in image-guided laparoscopic rectal surgery: results of a phantom study. _Surg. Endosc._ 30, 495–503 (2016).
Article CAS Google Scholar * Jhaveri, K. S. & Hosseini-Nik, H. MRI of rectal cancer: an overview and update on recent advances. _Am. J. Roentgenol._ 205, W42–W55 (2015). Article
Google Scholar * Li, X.-T., Zhang, X.-Y., Sun, Y.-S., Tang, L. & Cao, K. Evaluating rectal tumor staging with magnetic resonance imaging, computed tomography, and endoluminal
ultrasound: a meta-analysis. _Medicine (Baltimore)_ 95, e5333–e5333 (2016). Article Google Scholar * Kim, S. H., Lee, J. Y., Lee, J. M., Han, J. K. & Choi, B. I. J. E. R. Apparent
diffusion coefficient for evaluating tumour response to neoadjuvant chemoradiation therapy for locally advanced rectal cancer. _Eur. Radiol._ 21, 987–995 (2011). Article Google Scholar *
van der Paardt, M. P., Zagers, M. B., Beets-Tan, R. G., Stoker, J. & Bipat, S. Patients who undergo preoperative chemoradiotherapy for locally advanced rectal cancer restaged by using
diagnostic MR imaging: a systematic review and meta-analysis. _Radiology_ 269, 101–112 (2013). Article Google Scholar * Franz, A. M. et al. Electromagnetic tracking in medicine—a review of
technology, validation, and applications. _IEEE Trans. Med. Imaging_ 33, 1702–1725 (2014). Article Google Scholar * Janssen, N. et al. Real-time wireless tumor tracking during breast
conserving surgery. _Int. J. Comput. Assist. Radiol. Surg._ 13, 531–539 (2018). Article Google Scholar * Willoughby, T. R. et al. Target localization and real-time tracking using the
Calypso 4D localization system in patients with localized prostate cancer. _Int. J. Radiat. Oncol. Biol. Phys._ 65, 528–534 (2006). Article Google Scholar * Nijkamp, J. et al. Comparing
position and orientation accuracy of different electromagnetic sensors for tracking during interventions. _Int. J. Comput. Assist. Radiol. Surg._ 11, 1487–1498 (2016). Article Google
Scholar * Moningi, S. et al. Analysis of fiducials implanted during EUS for patients with localized rectal cancer receiving high-dose rate endorectal brachytherapy. _Gastrointest. Endosc._
81, 765–769.e761 (2015). Article Google Scholar * van den Ende, R. P. J. et al. MRI visibility of gold fiducial markers for image-guided radiotherapy of rectal cancer. _Radiother. Oncol._
132, 93–99 (2019). Article Google Scholar * Vorwerk, H. et al. Gold markers for tumor localization and target volume delineation in radiotherapy for rectal cancer. _Strahlentherapie und
Onkol._ 185, 127–133 (2009). Article Google Scholar * Campa-Thompson, M., Weir, R., Calcetera, N., Quirke, P. & Carmack, S. Pathologic processing of the total mesorectal excision.
_Clin. Colon Rectal Surg._ 28, 43–52 (2015). Article Google Scholar Download references ACKNOWLEDGEMENTS We would like to thank Koningin Wilhelmina Fonds—Alpe d’HuZes (NKI 2014-6596) for
their funding. AUTHOR INFORMATION Author notes * These authors contributed equally: Jasper Nijkamp, Theo J. M. Ruers. AUTHORS AND AFFILIATIONS * Department of Surgical Oncology, The
Netherlands Cancer Institute, Amsterdam, The Netherlands Esther N. D. Kok, Roeland Eppenga, Koert F. D. Kuhlmann, Harald C. Groen, Ruben van Veen, Wouter J. Heerink, Nikie J. Hoetjes,
Oleksandra Ivashchenko, Geerard L. Beets, Arend G. J. Aalbers, Jasper Nijkamp & Theo J. M. Ruers * Department of Gastrointestinal Oncology, The Netherlands Cancer Institute, Amsterdam,
The Netherlands Jolanda M. van Dieren, Thomas R. de Wijkerslooth & Monique van Leerdam * Department of Radiology, The Netherlands Cancer Institute, Amsterdam, The Netherlands Doenja M.
J. Lambregts * Faculty TNW, Group Nanobiophysics, Twente University, Enschede, 7522 NB, The Netherlands Theo J. M. Ruers Authors * Esther N. D. Kok View author publications You can also
search for this author inPubMed Google Scholar * Roeland Eppenga View author publications You can also search for this author inPubMed Google Scholar * Koert F. D. Kuhlmann View author
publications You can also search for this author inPubMed Google Scholar * Harald C. Groen View author publications You can also search for this author inPubMed Google Scholar * Ruben van
Veen View author publications You can also search for this author inPubMed Google Scholar * Jolanda M. van Dieren View author publications You can also search for this author inPubMed Google
Scholar * Thomas R. de Wijkerslooth View author publications You can also search for this author inPubMed Google Scholar * Monique van Leerdam View author publications You can also search
for this author inPubMed Google Scholar * Doenja M. J. Lambregts View author publications You can also search for this author inPubMed Google Scholar * Wouter J. Heerink View author
publications You can also search for this author inPubMed Google Scholar * Nikie J. Hoetjes View author publications You can also search for this author inPubMed Google Scholar * Oleksandra
Ivashchenko View author publications You can also search for this author inPubMed Google Scholar * Geerard L. Beets View author publications You can also search for this author inPubMed
Google Scholar * Arend G. J. Aalbers View author publications You can also search for this author inPubMed Google Scholar * Jasper Nijkamp View author publications You can also search for
this author inPubMed Google Scholar * Theo J. M. Ruers View author publications You can also search for this author inPubMed Google Scholar CONTRIBUTIONS K.F.D.K., M.v.L., O.I., J.N., and
T.J.M.R. contributed to the conception and design of the study. E.N.D.K. and N.J.H. included all patients. The implementation team including H.C.G., R.v.V., N.J.H., and W.J.H. performed all
clinical and technical proceedings concerning the navigation system before and during surgery. M.v.L., J.M.v.D., and T.R.d.W. performed all fiducial marker placements. D.M.J.L. and J.N.
validated all segmentations of the 3D models. G.L.B. and A.G.J.A. performed the surgical procedures. E.N.D.K. and R.E. executed the clinical study, analyzed the data, and wrote the
manuscript with assistance from all contributing authors. All authors critically revised the manuscript. CORRESPONDING AUTHOR Correspondence to Esther N. D. Kok. ETHICS DECLARATIONS
COMPETING INTERESTS The authors declare no competing interests. ADDITIONAL INFORMATION PUBLISHER’S NOTE Springer Nature remains neutral with regard to jurisdictional claims in published maps
and institutional affiliations. SUPPLEMENTARY INFORMATION NR-REPORTING-SUMMARY RIGHTS AND PERMISSIONS OPEN ACCESS This article is licensed under a Creative Commons Attribution 4.0
International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the
source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative
Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by
statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit
http://creativecommons.org/licenses/by/4.0/. Reprints and permissions ABOUT THIS ARTICLE CITE THIS ARTICLE Kok, E.N.D., Eppenga, R., Kuhlmann, K.F.D. _et al._ Accurate surgical navigation
with real-time tumor tracking in cancer surgery. _npj Precis. Onc._ 4, 8 (2020). https://doi.org/10.1038/s41698-020-0115-0 Download citation * Received: 09 January 2020 * Accepted: 12 March
2020 * Published: 08 April 2020 * DOI: https://doi.org/10.1038/s41698-020-0115-0 SHARE THIS ARTICLE Anyone you share the following link with will be able to read this content: Get shareable
link Sorry, a shareable link is not currently available for this article. Copy to clipboard Provided by the Springer Nature SharedIt content-sharing initiative