ABSTRACT Commercial wearable biosignal sensing technologies encounter challenges associated with irritation or discomfort caused by unwanted objects in direct contact with the skin, which
can discourage the widespread adoption of wearable devices. To address this issue, we propose a fabric-based lamina emergent MXene-based electrode, a lightweight and flexible shape-morphing
wearable bioelectrode. This work offers an innovative approach to biosignal sensing by harnessing the high electrical conductivity and low skin-to-electrode contact impedance of MXene-based
dry electrodes. Its design, inspired by Nesler’s pneumatic interference actuator, ensures stable skin-to-electrode contact, enabling robust biosignal detection in diverse situations.
Extensive research is conducted on key design parameters, such as the width and number of multiple semicircular legs, the radius of the anchoring frame, and pneumatic pressure, to
accommodate a wide range of applications. Furthermore, a real-time wireless electrophysiological monitoring system has been developed, with a signal-to-noise ratio and accuracy comparable to
those of commercial bioelectrodes. This work excels in recognizing various hand gestures through a convolutional neural network, ultimately introducing a shape-morphing electrode that
provides reliable, high-performance biosignal sensing for dynamic users. SIMILAR CONTENT BEING VIEWED BY OTHERS ADHESIVE FREE, CONFORMABLE AND WASHABLE CARBON NANOTUBE FABRIC ELECTRODES FOR
BIOSENSING Article Open access 09 December 2022 ALL-PRINTED NANOMEMBRANE WIRELESS BIOELECTRONICS USING A BIOCOMPATIBLE SOLDERABLE GRAPHENE FOR MULTIMODAL HUMAN-MACHINE INTERFACES Article
Open access 10 July 2020 3D PRINTABLE AND MYOELECTRICALLY SENSITIVE HYDROGEL FOR SMART PROSTHETIC HAND CONTROL Article Open access 21 January 2025 INTRODUCTION Wearable technology is
substantially gaining importance in the healthcare sector owing to various societal concerns, such as aging populations and the emergence of infectious diseases. Preventive medicine has the
potential to reduce social costs significantly by enabling early interventions against acute diseases1. However, the effective implementation of preventive medicine faces several challenges,
including difficulties in miniaturizing biodata acquisition systems, ensuring stable skin-to-bioelectrode contact, allowing continuous wear, and facilitating data analysis2. To address
these challenges, wearable healthcare systems have emerged as prominent candidates that can seamlessly collect and track user biodata3,4. Among the wearable healthcare systems,
tattoo-based5,6,7, and textile-based8,9 wearables have been extensively studied because of their potential for daily analysis of physiological diseases, such as cardiovascular diseases10,
muscle disorders11, Parkinson’s disease12, and Alzheimer’s disease13. Recent advancements in stretchable and flexible dry electrodes, which comprise one-dimensional (1D) or two-dimensional
(2D) conductive materials, have significantly improved their sensitivity, particularly for body suit-type applications14,15. Despite their advantages, wearable devices have persistently
caused issues of discomfort due to direct contact with the skin. Tokihiko et al. highlighted the importance of comfort-of-wear as a critical parameter for the future of wearables, similar to
selecting comfortable clothing16. This term describes the ability of a device to seamlessly integrate into a user’s daily life without causing any disruption or sense of irregularity
because of its physical presence. Although the size of current rigid wearable devices, such as wristwatches or rings, has been reduced to enhance comfort, the discomfort associated with
wearing such devices has failed to be eliminated16. Consequently, individuals sensitive to discomfort caused by irregular objects are discouraged from using these devices, which in turn
deters the adoption of wearable healthcare devices and limits the effectiveness of preventive medicine efforts. To tackle this issue, we present a fabric-based lamina emergent MXene-based
electrode (FLEXER) as a wearable biosensing electrode. This shape-morphing bioelectrode utilizes a laminar emergent mechanism crafted from thermoplastic polyurethane (TPU)-laminated cotton
fabric. The FLEXER has a symmetric whirlpool-shaped configuration that seamlessly transforms from a flat state to a scissor jack-like shape when pneumatic pressure is applied. The expanded
air pouch self-intersects, forming millimeter-sized joints that can be individually folded and unfolded, enabling conformal and stable electric contact with the skin even during motion. The
contact force of the FLEXER can be adjusted by controlling the air pressure and geometric characteristics of the joints. Additionally, the primary electrode material of the FLEXER is MXene,
a 2D material comprising transition metal and carbon with excellent electrical conductivity and mechanical durability17. To further enhance the mechanical properties18 and oxidation
resistance19,20 of MXene, it was supplemented with cellulose nanofiber (CNF)21 and polycarboxylate ether (PCE)22. This innovative FLEXER technology can seamlessly integrate into everyday
clothing, allowing for noninvasive monitoring of electrocardiogram (ECG) and electromyogram (EMG) signals. By combining the advantages of pneumatic lamina emergent systems and flexible MXene
electrodes, we were able to develop a concealable, fast, safe, and reliable biosensing electrode that revolutionizes the field of wearable health monitoring. RESULTS CONCEPT OF FLEXER AS A
WEARABLE DEVICE The FLEXER-based wearable healthcare system differs from conventional wearable healthcare systems in that it has a concealable design owing to its shape-morphing capability
(Fig. 1a (left)). Unlike other wearable systems that use bioelectrodes requiring constant direct skin contact, the FLEXER eliminates the constant need for physical contact. Traditional
bioelectrodes, particularly dry ones, require pressure application, which is usually accomplished with biomedical adhesives or pressure bandages23. However, the unique design of the FLEXER
allows it to be activated or deactivated as required (see Fig. 1a (right)), providing a user-friendly and adaptable solution. The operation of the FLEXER is elucidated by Nesler’s pneumatic
interference actuator (PIA) model24 (Fig. 1b). The PIA model can generate large deformation by strategically using the flexible structure’s crease. Specifically, when the flexible air pocket
is folded, it forms an overlapping volume at the creased section (Fig. 1b, State I). The overlapped volume decreases the overall volume of the air pocket below the original volume.
Consequently, the air pocket tends to revert to its original volume by flattening the crease (i.e., unfolding) when subjected to pneumatic pressure (Fig. 1b, State II). During this process,
the crease operates as a torque pivot point, inducing significant deformation in the structure24,25. The shape-morphing behavior of the FLEXER can be precisely controlled by coordinating
multiple PIA joints. The symmetrical arrangement of double-folded PIA joints, with one end touching the anchoring frame and the other end touching the center, enables the controlled movement
of the center electrode (CE) along an axis perpendicular to the contact plane of the FLEXER (Fig. 1c). The FLEXER system comprises three key components, as depicted in Fig. 1d. The first
component is the MXene/CNF/PCE (MX-CNF-PCE) composite electrode with a high electrical conductivity, flexibility, and oxidation resistance, and it is placed at the upper skin contact area.
This electrode is seamlessly integrated into cotton fabric using a solution spray method. It leverages the intricate 3D woven structure of the fabric to form a conductive network of MXene in
a 3D configuration. The 3D MXene electrode structure effectively confines mechanical damage to the outermost surface, thereby aiding in the preservation of conductivity in the event of
electrode delamination due to adhesion failure26. (Supplementary Fig. 1) The second component utilizes TPU, which is known for its mechanical resilience and stretchability, to create an air
pocket that is a fundamental part of the PIA system within the FLEXER27. TPU films have a low melting point of 140–150 °C, making them easy to meld28 using a customized XY plotter with a
soldering iron. The TPU-laminated cotton fabric was precisely cut using a customized computer numerical control (CNC) cutting machine (Fig. 1e and Supplementary Figs. 2 and 3). Finally, the
outer layer of the FLEXER is made from cotton fabric, which serves as its sheath. When users wear clothing embedded with the FLEXER, the overall appearance of the attire closely resembles
regular clothing (Supplementary Fig. 4). This distinguishes it from other wearable devices that usually have a tighter fit29. Notably, the FLEXER remains inconspicuous externally,
effectively reducing the possibility of social stigma associated with wearing it. The fabricated FLEXER was connected to an airline from a handheld air blower and stitched into the clothing
(Fig. 1f). The advantages of the FLEXER also include structural simplicity, which ensures durability without intricate mechanical component requirements; rapid actuation due to friction-free
pneumatic pressure; and paramount user safety, which stems from the absence of rigid parts and its PIA-based mechanism that allows for low pneumatic pressure usage and minimizes potential
risks. PHYSICAL AND DESIGN ANALYSIS OF FLEXER The straightforward linear structure with a double-fold design was examined to comprehensively explore the relationship between the pneumatic
structure of the FLEXER and its performance and design parameters. The linear PIA, measuring 11.8 cm in length and 1.9 cm in width (see Fig. 2a, top photographs), underwent an unfolding
actuation, which was recorded and analyzed using video analysis software (Kinovea). The inherent material characteristics of the cotton fabric caused the linear PIA to exhibit a
self-unfolding tendency even without the application of air pressure, as illustrated in Supplementary Fig. 5. To counteract this spontaneous unfolding, rubber magnets were affixed near the
linear PIA joints using magnetic force to impede undesired expansion (see Supplementary Fig. 6). Thereafter, the movement of the linear PIA end point and the angle changes of its joints were
observed during actuation. The graph in Fig. 2a (bottom right) illustrates that Joint 1 underwent rapid inflation and expansion between 100 and 125 ms, ultimately reaching a 180° angle by
175 ms (Supplementary Movie 1). Simultaneously, Joint 2 gradually expanded between 100 and 250 ms and then aggressively expanded at 250 ms after Joint 1 reached a straight angle. This
behavior is probably due to the folded portions of the linear PIA weakly obstructing airflow, leading to a sequential inflation process. By 375 ms, all the joints had fully expanded to their
maximum distance of 7.55 cm. Although the length of the linear PIA decreased from its initial value (11.8 cm) because of perpendicular expansion, the overall change in dimensions remained
minimal (see Supplementary Fig. 7 and Supplementary Movie 2 for further clarification). The linear PIA actuation behavior revealed that the joint close to the air pressure application point
tends to actuate faster than the joint at a further distance. This occurs because of the 180° bending angle of the airflow path that creates large drops of air velocity. Since the FLEXER has
multiple semicircular legs that require synchronized actuation, the parts that transfer air pressure to each leg were designed in a circular shape to minimize airflow interference when air
pressure is applied. As discussed earlier, the folded self-intersecting sections within the air pouch joints function as springs, directing the unconstrained end of the air pouch to move in
a manner that minimizes the volume of self-intersection. However, the actuation of a single-block air pouch actuator involves both lateral and vertical movements, making it difficult to
precisely control the direction of desired inflation. This limitation necessitates the incorporation of multiple joints to effectively inhibit the lateral actuation of the air pouch. Our
devised FLEXER addresses this necessity through a device architecture that encompasses (i) a CE that serves as an electric contact point, (ii) an anchoring frame (AF) that maintains the
shape of the FLEXER and provides the necessary mechanical constraint for actuation, and (iii) multiple semicircular legs (SLs) strategically positioned in an equiangular arrangement to link
the AF and the CE (Supplementary Figs. 8 and 9). As depicted in the top photographs in Fig. 2b, a three-legged FLEXER with a 10 cm AF diameter was fabricated to illustrate the design
approach. The AF is securely affixed to the fabric, serving as an integral mechanical anchor for the entire structure. The dimensions of the CE can be flexibly adjusted by either being
enlarged to enhance signal detection or reduced to minimize the overall size of the FLEXER. The SLs are linked to both the CE and AF through two pivotal joints. The first joint, which is
situated at the AF, dictates the rotational direction of the SLs while functioning as a stabilizer at the CE level. Similarly, the second joint, which is located at the CE, serves as an
electrode-level stabilizer, ensuring the stability of the entire system. When pneumatic pressure was applied, both joints unfolded, directing the CE vertically. While expansion, FLEXER’s
Joint 1 raised the SLs and CE by deflecting it from 0° to 42.2° (Fig. 2b, bottom right graph), and Joint 2 shifted the angle from 0° to 31.5°, ensuring parallel alignment between the CE and
AF planes for stable skin contact. The FLEXER also achieved a rapid actuation speed of ~12 cm/s by deploying the CE in <0.4 s (Fig. 2b, bottom left graph). This rapid actuation proves
advantageous in situations30 where users require prompt analysis of vital signals (e.g., ECG and EMG signals) related to urgent medical conditions, such as coronary heart disease,
cardiovascular disease, and congestive heart failure31,32,33,34. Furthermore, we assessed the contact stability of the FLEXER during its spring-like motion. The three-legged FLEXER was
attached to a linear actuator, and a cyclic linear motion with a 20 mm travel length was applied to simulate dynamic movement while measuring the contact force of the CE. As a comparative
measure, a balloon with the same diameter as the AF of the FLEXER was also tested (Fig. 2c and Supplementary Movies 3 and 4). The measured force was then differentiated with respect to time
to quantify the change in force over time. The result revealed that the three-legged FLEXER exhibited a consistent time derivative of force during repetitive movement, highlighting its
effectiveness in reducing motion-related interference during biosensing35. Contrastingly, the balloon displayed a standard deviation of the time derivative of force of 0.212 N/s, which is 80
times higher than that of the FLEXER (0.0026 N/s). This difference underscores the enhanced contact stability of the FLEXER in managing movement-related challenges during sensing
operations. Moreover, we evaluated the deviation of the CE from its center after the activation of FLEXER because MXene was coated on only one SL of the FLEXER, extending from the SL to the
AF as a conductive trace. Supplementary Figure 10 presents top-view images comparing the activation of MXene-coated and non-coated FLEXER samples. The CE diameter of the FLEXER is 10 mm, and
the center of the CE remained aligned with the inner contact area after activation. This demonstrates the consistent movement of the FLEXER system along with the electrode materials. To
expand the FLEXER into a wider range of wearable bioelectrodes, the contact force of the electrodes must be adjustable. This can be achieved by selecting an appropriate shape and number of
SLs. To evaluate the relationship between the FLEXER design parameters and the resulting contact force, a set of 16 distinct samples with varying SL widths and numbers were fabricated (Fig.
2d, e). Since the AF remained mechanically fixed and the movement of the CE depended on the behavior of the SLs, the design parameters for the FLEXER were confined to the shape and number of
the SLs. The AF and CE diameters were maintained at 130 and 15 mm, respectively, while the pneumatic pressure was maintained at 5 psi to eliminate any additional control variable. Figure 2f
presents the recorded contact forces of the different SL widths and numbers for the FLEXER. The data demonstrate that the contact force increased as the number and width of the SLs
increased. This trend can be explained by the PIA mechanism, which establishes a connection between the output force and the volume of joint overlap. Regarding the FLEXER, the increase in
the number and width of its SLs directly contributes to the enlargement of the overlapping areas. This increase in overlapping areas leads to a heightened contact force. Eq. (1) can be used
to determine the contact force based on the two design parameters at a pneumatic pressure of 5 psi: $$F=0.25+0.35 * w+0.24 * n \, ({{{R}}}^{2}=0.97),$$ (1) where _F_ represents the contact
force, and _w_ and _n_ represent the width and number of the SLs, respectively. This empirical relationship is graphically illustrated in Fig. 2f. The contact force is expressed as follows
at high pneumatic pressures of 10 psi (Eq. (2)) and 15 psi (Eq. (3)) (Supplementary Fig. 11): $$F=0.26+0.33 * w+0.30 * n \, ({{{R}}}^{2}=0.93),$$ (2) $$F=0.28+0.32 * w+0.34 * n \,
({{{R}}}^{2}=0.90)$$ (3) Additionally, the maximum contact distance of the FLEXER electrode depends on the FLEXER size, as shown in Supplementary Fig. 12. When the AF diameter of the FLEXER
varied from 8 to 13 cm, the contact distance increased from 1.7 to 3.2 cm, all at a constant pressure of 10 psi. Considering both the magnitude and uniformity of the contact force is
essential for the FLEXER to effectively function as a bioelectrode in accurately detecting biological signals. To assess the impact of contact force uniformity based on the number of SLs,
the contact force of the CE was measured at various points using a 3 × 3 grid, as depicted in the schematic at the top-left corner of Fig. 2g. The contact force distributions of FLEXER
models with two, three, four, and five SLs were analyzed, as shown in Fig. 2g and Supplementary Fig. 13. Calculating the standard deviation for FLEXER models with different numbers of SLs
revealed the following trend: An increase in the number of SLs resulted in a more uniform distribution of contact force, as illustrated in Fig. 2h. CHEMICAL ANALYSIS OF MX-CNF-PCE ELECTRODES
MXene was used as the electrode material of the FLEXER because of its exceptional electrical conductivity and compatibility with water-based solution methods. However, MXene has poor
mechanical strength and experiences a reduction in electrical conductivity under ambient oxidation. To address these challenges, we incorporated PCE and CNF into the MXene solution to
enhance the solution properties. When combined with MXene, PCE effectively envelops the MXene surface, providing a protective shield against oxygen exposure. Moreover, this combination
enhances the flexibility of MXene by widening the gaps between its platelets. Conversely, CNF significantly increases the mesoscopic mechanical strength of the composite. Since MXene has a
platelet-like structure, the interconnection of individual MXene platelets through covalent bonds can result in poor mechanical strength. CNF serve as a platform for MXene attachment because
their dimensions (diameter: 3–30 nm and length: several microns) differ significantly from those of MXene (diameter: several microns and thickness: 1.2–1.8 nm36), facilitating the release
of strain within the composite structure. Supplementary Figure 14 provides a schematic representation of this composite, which encompasses both CNF and PCE elements. To realize this
approach, a water-based solution containing MXene, CNF, and PCE (MX-CNF-PCE) was prepared. This MX-CNF-PCE solution was then spray-coated onto the contact side of the FLEXER structure using
a polyethylene terephthalate (PET) shadow mask (Fig. 3a and Supplementary Fig. 15). The hydrophilic nature of the cotton fabric, the hydrophilic functional groups present on the MXene
nanosheets, and the hydroxyl groups within the CNF and PCE all enabled the solution to be adsorbed onto the fabric, resulting in the formation of densely packed MX-CNF-PCE layers. We
confirmed the uniformity and integrity of the MX-CNF-PCE spray-coating using scanning electron microscopy (SEM) and energy-dispersive X-ray (EDX) analysis37, as illustrated in Fig. 3b. The
top-view SEM image reveals that the MXene nanosheets enveloped the cotton fibers and bridged the gaps between them, while the unmodified cotton fabric exhibited a clean surface characterized
by interwoven warp and weft yarns (Supplementary Fig. 16). Additionally, the uniform distribution of each element within the MXene-based solution was observed from the EDX mapping (Fig. 3b
(right side) and Supplementary Fig. 17). Generally, electrodes must have low sheet resistance to be able to measure biosignals with high reliability. Here, the MX-CNF-PCE and MX-CNF
electrodes were measured to have sheet resistances of 24.7 Ω/☐ and 20.5 Ω/☐, respectively, which were slightly higher than that of the MXene electrode (19.1 Ω/☐) (Fig. 3c). This is
attributed to the inclusion of electrically nonconductive CNF and PCE within the MXene interlayers. However, their sheet resistances are still significantly lower than those of previously
reported MXene-coated cotton electrodes (569–761 Ω/☐)37. In addition to the sheet resistance of electrodes, the skin-to-electrode impedance is also a crucial factor. Skin-to-electrode
impedance reflects the skin’s resistance in transmitting electrical signals from the user to the sensing element of the electrode. Low skin-to-electrode impedance is associated with high
biosignal quality, as it signifies reduced resistance at the contact point between the skin and conductive surfaces38. Fig. 3d illustrates the measured skin-to-electrode contact impedances
of the FLEXER electrodes and reference dry and gel-type Ag/AgCl electrodes. The inflated FLEXER electrode displayed skin-to-electrode impedance comparable to those of reference electrodes of
the same size (diameter = 3 cm). Notably, the skin-to-electrode impedance of the FLEXER electrode was similar to that of the dry-type electrode, particularly up to a frequency of 2.45 kHz.
At a frequency of 100 Hz, the FLEXER electrode exhibited an impedance of 191 kΩ, while the dry and gel Ag/AgCl electrodes exhibited impedances of 190 kΩ and 64 kΩ, respectively39. These
impedance values are sufficiently low for accurate biosignal measurements without significant noise interference40. Further, we investigated the electromechanical responses of the MX-CNF-PCE
bioelectrode under mechanical deformation. A 15% tensile strain was applied to the three electrodes (MXene, MX-CNF, and MX-CNF-PCE), and the normalized resistance (R/R0) was measured, as
depicted in Fig. 3e. MX-CNF-PCE showed the best electromechanical stability with the smallest R/R0 increase of 2.1 at a 15% tensile strain, whereas MXene and MX-CNF exhibited R/R0 values of
4.8 and 3.2, respectively. The enhanced stability of MX-CNF-PCE is attributed to its unique brick-and-mortar architecture. In this structure, the 2D MXene nanosheets function as bricks,
while the 1D CNF serves as the mortar41. This is due to the formation of hydrogen bonds between the terminal groups (–OH, –O, and –F) of MXene and the hydroxyl groups of the CNF. The
incorporation of PCE, a comb-type polymer, into MX-CNF-PCE introduces additional hydrogen bonds, which significantly enhance its mechanical stability. Owing to these improvements, MX-CNF-PCE
exhibits a decrease in R/R0 in stark contrast to MX-CNF. Furthermore, we conducted a cyclic bending stability test by subjecting the three MXene-based electrodes to a 15% strain repeatedly
for 1000 cycles while measuring the R/R0 values, as depicted in Fig. 3f. This test simulates situations where stress is frequently applied to the bioelectrode on the fabric overlying
specific areas, such as joints like knees or elbows42. A consistent trend was observed throughout the test: The inclusion of CNF and PCE enhanced electromechanical stability. MX-CNF-PCE,
MX-CNF, and MXene had R/R0 values of 1.4–1.6, 2.3–2.6, and 6.1–16.9, respectively. For the oxidation resistance test, the MXene-based electrodes were exposed to a sealed glass chamber with
100% relative humidity for 17 days, as illustrated in Supplementary Fig. 18. The results of this rigorous environmental test revealed the remarkable oxidation stability of MX-CNF-PCE, as
evidenced by a modest increase in its R/R0 value from 1 to 65 (Fig. 3g). This marginal increase in R/R0 became apparent when compared to those of the MXene and MX-CNF electrodes43, which
experienced substantial increases to 1231 and 548 (20 times and 9 times that of MX-CNF-PCE), respectively. As mentioned earlier, this is attributed to the protective barrier formed by the
incorporation of PCE, which effectively shields MXene from oxidation. Sweat excreted by the skin can also affect the performance of bioelectrodes, thereby affecting their R/R0 ratio. To
assess the stability of MX-CNF-PCE in response to perspiration, its electrical characteristics were observed in an artificial sweat44. After wetting the three MXene-based electrodes with the
artificial sweat (see Supplementary Fig. 19), we monitored their electrical response (R/R0) for 20 minutes (Fig. 3h). MX-CNF-PCE exhibited the most electrically stable response because of
the PCE coating on the MXene nanosheets. Additionally, experiments were conducted to determine the optimal mixing ratio of MX-CNF-PCE for enhanced electromechanical properties by doubling
and halving the CNF and PCE amounts, as shown in Supplementary Fig. 20. Since CNF and PCE act as insulators for the electrode, their proportion needs to be minimized in comparison to MXene.
Considering the spray-coating capability and the electrical conductivity of the bioelectrode, we determined the optimized mass ratio of MXene to be 82 wt% of the total electrode composition.
Maintaining this MXene weight ratio (82 wt%), we precisely adjusted the amounts of CNF and PCE in the electrode to further improve its environmental stability. By comparing the
environmental stability of bioelectrodes with different CNF and PCE ratios, we identified the weight ratio of 24:4:1 (MXene:CNF:PCE) as optimal for the FLEXER system. Consequently, the
MX-CNF-PCE electrode demonstrated enhanced electromechanical and environmental resistance properties under various mechanical deformations, including tensile strain and cyclic bending, 100%
humidity environment, and interference from sweat. Spectroscopic characterization was performed on the MX-CNF-PCE electrode to understand the chemical structure underlying its improved
electromechanical test results (Fig. 3i‒k). The X-ray diffraction (XRD) curves in Fig. 3i and Supplementary Fig. 21 reveal evidence of the intercalation and blending of the CNF and PCE
within the MXene nanosheets. Specifically, the (002) diffraction peak for MXene appeared at 2θ = 5.82°, which shifted to 5.70° in MX-CNF and 4.66° in MX-CNF-PCE. These shifts correspond to
an increase in the interlayer spacing from 3.04 nm to 3.10 nm and 3.78 nm, respectively22,45,46,47. Fig. 3j presents the X-ray photoelectron spectroscopy (XPS) spectra of the pristine MXene,
MX-CNF, and MX-CNF-PCE composite films. An analysis of the Ti 2_p_ and O 1 _s_ spectra reveals that the area ratio of Ti2+ to Ti3+ peaks increased from 3.52 (MXene) to 3.56 and 3.85 (MX-CNF
and MX-CNF-PCE, respectively). Simultaneously, the area ratio of the –COO– peak to the total chemical bonds decreased from 0.04 (MXene) to 0.02 (MX-CNF and MX-CNF-PCE). These observations
confirm the formation of covalent Ti–O bonds between the MXene nanosheets and the PAA anchor of PCE, as revealed by the Ti 2_p_ spectrum (i.e., –COOH). Furthermore, the F 1 _s_ spectra
reveal the presence of hydrogen bonds between the MXene nanosheets and the hydroxyl groups in the CNF or the partially hydrolyzed –COO– of PCE (i.e., –COOH). The area ratio of the hydrogen
bond peak (C–Ti–Fx –H) to the total area increased from 0 (MXene) to 0.2 and 0.32 (MX-CNF and MX-CNF-PCE, respectively). The appearance of a high binding energy peak in the F 1 _s_ region
for C–Ti–Fx–H demonstrates the formation of hydrogen bonds, which contributed to MX-CNF and MX-CNF-PCE exhibiting better mechanical stability than pristine MXene. These ratio increments are
depicted in Fig. 3k. WIRELESS ELECTROPHYSIOLOGICAL MONITORING USING FLEXER To showcase the capabilities of the FLEXER in wearable biosensing, we developed a wireless electrophysiological
monitoring system. Figure 4a provides an overview of the working concept and block diagram of the developed system (see Supplementary Figs. 22 and 23 for more details). The FLEXER and the
flexible printed circuit board (fPCB) were electrically connected through jumper wires, allowing electrophysiological signals detected by the FLEXER to be routed to the biopotential
amplifier on the fPCB. These signals were then amplified and filtered by the biopotential amplifier before undergoing analog-to-digital conversion (ADC). The converted signals were
transmitted wirelessly to the computing unit in real time via a Bluetooth Low-Energy (BLE) module for further analysis. An ECG measurement was conducted with the developed wireless
electrophysiological monitoring system (Fig. 4b, see Supplementary Fig. 24 left for details). As a control group, ECG signals were also measured using commercially available gel-type and dry
Ag/AgCl electrodes with the developed wireless system. To record the ECG signals, two electrodes were placed on the left and right sides of the chest, and a ground electrode was placed on
the lower abdomen48. During the ECG recording, the FLEXER was affixed to the clothing, while both gel-type and dry Ag/AgCl electrodes were securely attached to the skin. Despite the gap
between the clothing and the skin, the FLEXER exhibited a high signal-to-noise ratio (SNR) because of its unique pneumatic structure, which facilitated effective skin-electrode contact (Fig.
4c). Conversely, the FLEXER with low air pressure displayed a low SNR owing to inadequate skin-electrode contact. Consequently, the PQRST waveforms in the ECG recorded with the FLEXER
closely resembled those obtained using commercial Ag/AgCl gel electrodes (Supplementary Fig. 24 right). Moreover, the FLEXER demonstrated its efficacy in monitoring changes in heart rate via
ECG measurements performed before (72 BPM) and after (125 BPM) exercise, as shown in Fig. 4d, e. Unlike commercial Ag/AgCl gel electrodes, which necessitate continuous skin contact for
long-term monitoring, FLEXERs only make contact with the skin during measurements. This minimizes discomfort even during extended periods of use, which is crucial for ensuring user comfort
and compliance in continuous health monitoring scenarios. Supplementary Movie 5 shows emergent status of single-FLEXER and multi-FLEXER when pneumatic force is occurred which can contact to
skin with facile hand pumping. Real-time measurement of ECG using multi-FLEXER in situations involving standing, walking, and driving, both in the inactive and active status has been also
proceeded in Fig. 4f. During real-time ECG monitoring, signals were dramatically got stabilized after the activation of FLEXER. (Supplementary Movie 6) The FLEXER is also highly suitable for
capturing EMG signals, which are foundational for hand gesture recognition, because of its enhanced electrical conductivity and stable skin contact due to its pneumatic-based design. This
capability has profound implications in areas such as human-machine interfaces and particularly prosthesis control in healthcare45,46,49. Gesture recognition via the EMG process entails the
placement of multichannel electrodes on various areas of the arm, with recognition being facilitated by extracting EMG signal features using various classifiers (Fig. 5a)47. To demonstrate
this, EMG signals corresponding to eight distinct gestures were measured using FLEXERs positioned on the forearm, as depicted in Fig. 5b. The recorded EMG signals, captured across five
channels for each gesture, displayed distinct patterns among the gestures. The signals were then subjected to the continuous wavelet transform (CWT), which transformed them into scalograms,
i.e., image-like time-frequency representations. These scalograms served as inputs for a multi-input convolutional neural network (CNN)-based gesture recognition system50. For each gesture,
the scalograms derived from the five channels underwent independent processing through three convolution layers and one fully connected layer to extract unique features. These processed
outputs were then concatenated to create a comprehensive feature set for each gesture. This concatenated layer was subsequently linked to three additional fully connected layers to classify
the eight gestures (Fig. 5c). Further details can be found in Supplementary Table 1 and Supplementary Figs. 25 and 26. To evaluate the generalizability of the system across independent
datasets, five-fold cross-validation tests were performed on randomly mixed datasets. The high-dimensional features extracted from the multi-input CNN model were visualized using
t-distributed stochastic neighbor embedding (t-SNE). This visualization displayed a distinct separation of features, contributing significantly to the achievement of an accuracy rate of
97.92% in Fold 5 (Fig. 5d and Supplementary Fig. 27). The recognition accuracy using CNN model for all eight gestures in each fold is presented as a confusion matrix (Fig. 5e and
Supplementary Fig. 28). The average accuracy across the five-fold cross-validation tests was 95.94%. This high accuracy is mainly due to the aforementioned SNR of the FLEXER. Generally, EMG
signals can be used to assess grip strength15,51. Thus, a grip strength test was conducted by performing gripping actions at various force levels (Fig. 5f). During this procedure, a
dynamometer quantified the force level, and the corresponding EMG signals were recorded using the FLEXER on the flexor carpi radialis of the forearm (Supplementary Fig. 29). The recorded EMG
signals were then segmented into two-second time windows, and the root mean square (RMS) of the signals were calculated. As depicted in Fig. 5g, the RMS of the EMG signals exhibited a
corresponding increase with grip strength. DISCUSSION In conclusion, the development of FLEXER, a lightweight and flexible wearable bioelectrode with a pneumatic structure, represents a
significant advancement in real-time healthcare signaling technology for wearable devices. The incorporation of MXene-based dry electrodes allows the FLEXER to achieve exceptional electrical
conductivity and significantly reduce skin-to-electrode contact impedance, thereby addressing key challenges in biosensing applications. Moreover, the innovative design of the FLEXER
ensures consistent skin contact, enabling stable signal recording and accurate monitoring of physiological parameters. Comprehensive evaluations demonstrated the impressive electromechanical
properties of the FLEXER, which include exceptional mechanical stability, strain resistance of up to 15%, and remarkable durability over 1000 bending cycles. Additionally, the real-time
wireless electrophysiological monitoring system developed using the FLEXER enables seamless recording and transmission of ECG and EMG signals. Furthermore, the FLEXER demonstrates
exceptional accuracy in monitoring curved regions, with a recognition accuracy of 95.94% for hand gesture recognition using EMG signals. These significant advancements highlight the
potential of FLEXERs for personalized healthcare applications, such as prosthesis control and rehabilitation, and underscore their transformative impact on the field of wearable
bioelectrodes. METHODS FABRICATION OF FLEXER The FLEXER design was simplified using the visual programming software Rhino 7 and Grasshopper 3D (Supplementary Fig. 9). Two pieces of
TPU-laminated cotton fabric were overlapped, ensuring that the TPU layers came into direct contact. Hot sealing with a pen-type soldering tip was conducted along the designed pathway using a
customized commercial XY plotter (Liyang equipment), and the fabric was cut to the desired size using Cameo 4 pro (Silhouette). After preparing the pneumatic structure of the FLEXER, the
MX-CNF-PCE solution was spray-coated on it through a PET shadow mask to create a conductive path on the FLEXER. Pneumatic force was applied to the FLEXER through an air blower connected to
an airline. SYNTHESIS OF TI3C2TX MXENE AND MXENE-BASED ELECTRODES The etchant solution was prepared in a 250 mL Teflon beaker as follows: 3.2 g of LiF (99.9%; Sigma-Aldrich) was dissolved in
40 mL of a 9 M HCl aqueous solution and cooled to <5 °C using an ice-water bath. The etchant was then deoxygenated with ultrahigh-purity N2. Subsequently, 2.0 g of MAX phase (Ti3AlC2
powder, 99%, 325 mesh; Forsman) was gradually added to the etchant under gentle stirring. The mixture was deoxygenated again and protected in an inert atmosphere. Etching was performed in an
oil bath at 25 °C for 48 h to remove most of the Al layer from the MAX phase. The resulting product was washed and delaminated in two 50 mL centrifuge tubes using deoxygenated water and
then subjected to five to six cycles of centrifuge-shaking (Labogene 416, ONE Commercial shaker Horizontal Single Head Shaking Machine 220 V LJS80-1). Finally, the delaminated MXene was
sonicated in an ice bath for 30 min and centrifuged at 2826×_g_ for 30 min to obtain a supernatant predominantly composed of single-layer MXene. The MXene dispersion was mixed with a CNF
polymer aqueous solution (1 wt%; CNNT) and a PCE polymer aqueous solution (MW = 50 k, 55.0 wt%; L’BESTE GAT) according to the mass ratio specified in the experimental design (MXene:CNF:PCE =
24:4:1, 24:2:1, 24:4:2, and 24:2:2). The mixture of MXene and polymers was vigorously shaken for 30 min to obtain a homogeneous MXene–polymer composite material. CHARACTERIZATION All video
analysis of the pneumatic structure of the FLEXER was performed using the Kinovea program. The contact force of the FLEXER was measured using a digital force gauge (ZTA-5N; IMADA) under
various pressure and FLEXER design conditions. The surface morphologies and elemental distribution of the MXene-based electrodes were characterized using SEM (JSM-7610F-Plus) and EDX
analyses. The sheet resistance was measured using the four-point probe method (Keithley 2182A, 2634B, and 6221). The skin-to-electrode impedance was measured using an
inductance–capacitance–resistance (LCR) meter (E-4980AL; KEYSIGHT) in the frequency range of 20 Hz to 300 kHz. Contact stability and tensile strain stability tests were conducted to evaluate
the mechanical robustness of the MXene-based electrodes on the FLEXER using a stepper motor controller (ECOPIA, BM-111). All electromechanical resistance data of the MXene-based electrode
were gathered using a parameter analyzer (4200A-SCS; Keithley). XRD (SmartLab; Rigaku) was used to detect the differences in the interlayer spacing of the MXene-based electrodes. An X-ray
photoelectron spectrometer (K-alpha, Thermo U.K.) was operated using an aluminum anode as the source with a spot size of 400 μm. WIRELESS ELECTROPHYSIOLOGICAL MONITORING An electrophysiology
amplifier (RHD 2132, Intan Technologies) with a 16-bit ADC and a differential gain of 192 was used to amplify, filter, and digitize electrophysiological signals. The sampling frequency was
258 Hz, and the bandpass filter frequency was 0.5–100 Hz. A BLE system on chip (QN9080SIP; NXP Semiconductors) wirelessly transmitted the data to a recording laptop for further analysis. The
entire system on the fPCB was powered by a 3.7 V lithium polymer battery and electrically connected to the FLEXER via jumper wires (AWG 30). REAL-TIME HEART RATE MEASUREMENT USING ECG The
ECG was measured by two electrodes placed on the chest and a ground electrode placed on the lower abdomen. In addition to the FLEXER, gel-type (Vitrode L-150X; Nihon Kohden) and dry (Laxtha)
Ag/AgCl electrodes were used for noise comparison of the signals. The Ag/AgCl dry electrodes were attached to the skin using medical tape (2702; 3 M). The commercial electrodes were
connected to the developed fPCB via snap connectors (Laxtha). The recorded ECG signals were transmitted wirelessly to the laptop through the BLE system on chip of the developed wireless
monitoring system. A fourth-order Butterworth band-pass filter (0.5–35 Hz) was used to denoise artifacts such as motion or power line artifacts. Subsequently, the R peak was detected to
calculate the heart rate in beats per minute. The SNR of a single PQRST wave in the ECG signals was calculated as follows: $${{{{{{\rm{SNR}}}}}}}_{{dB}}=20 \,{\log
}_{10}\frac{{A}_{{PQRST}}}{{A}_{{noise}}},$$ (4) where APQRST and Anoise are the amplitudes of the PQRST and noise, respectively. Thereafter, the SNR was averaged over the PQRST waves in the
recorded ECG signals. The signal processing was performed using MATLAB 2022b (MathWorks). The real-time ECG was measured with the subject standing, walking, and driving, in both the
inactive and active statuses of multi-FLEXER. GESTURE RECOGNITION AND GRIP STRENGTH TEST USING EMG For gesture recognition, five FLEXERs were placed on the forearm muscle. In addition, a
FLEXER was placed on the flexor carpi radialis for grip strength test. Both the ground and reference electrodes were positioned proximally to the elbow. Thereafter, the recorded EMG signals
were transmitted wirelessly to the laptop using the aforementioned system. Power line interference in the EMG signals was removed via a 60 Hz notch filter, and the signals were subsequently
filtered using the fourth-order Butterworth band-pass filter (10–80 Hz). Gesture recognition was conducted on 120 EMG signal datasets for each eight distinct gestures. In these datasets, the
signals were z-normalized and then segmented using a 4.5 s time window. Subsequently, scalograms were extracted through the CWT using the Morlet wavelet to provide image representations
that could serve as input for a multi-input CNN-based classifier. The scalogram image was then resized to 64 × 64 and normalized. The input size was set at 64 × 64 × 3 to account for the RGB
layers within each image. The architecture of the model comprised three convolution layers and one fully connected layer per channel, which were subsequently concatenated by a concatenation
layer. This concatenated layer was connected to three additional fully connected layers and was ultimately followed by a Softmax layer for the recognition of the eight gestures. A rectified
linear unit (ReLU) was selected as the activation function for the model. Meanwhile, the Adam optimizer (learning rate = 0.0001) was used as an optimizer, and loss consumption was performed
using the cross-entropy loss function. Five-fold cross-validation was performed for generalized performance. After training for the multi-input CNN for 100 epochs, the accuracy of the
five-fold cross-validation was averaged. The grip strength test was conducted using a dynamometer, with the volunteer maintaining a fixed elbow position. The dynamometer measured the maximum
force level as the volunteer applied various grip strengths with the right hand. EMG signals were simultaneously recorded by the FLEXER on the flexor carpi radialis. These EMG signals were
segmented using a 2 s time window to ensure that all grip-related signal variations were included. The RMS value of the segmented EMG signals was computed to establish a correlation between
the maximum force level and the RMS value derived from the grip-induced EMG signals. All signal processing was performed using MATLAB 2022b (MathWorks), and deep learning for gesture
recognition was performed using TensorFlow (Google). ETHICS INFORMATION FOR HUMAN SUBJECT RESEARCH The authors confirmed that under Article 13, Paragraph (1) of the Enforcement Rule of
Bioethics and Safety Act of the Ministry of Health and Welfare, Korea, Institutional Review Board approval was not needed as volunteers used wearable sensors with simple contact measurement.
In total of 2 participants were male and average age was 27 years old. The authors affirm that human research participants provided informed consents for publication of the images in Fig.
4f and Supplementary Figs. 4, 25, and 29. STATISTICS AND REPRODUCIBILITY The SEM images in Fig. 3b and Supplementary Figs. 16 and 17 were tested with each three samples and five times
repeated for each sample to get similar results. REPORTING SUMMARY Further information on research design is available in the Nature Portfolio Reporting Summary linked to this article. DATA
AVAILABILITY All data supporting the findings of this study are available within the article and its supplementary files. Any additional requests for information can be directed to, and will
be fulfilled by, the corresponding authors. Source data are provided with this paper. CODE AVAILABILITY The custom codes used in this study for wireless electrophysiology monitoring and
signal analysis are available from the corresponding author upon request. REFERENCES * Tricoli, A., Nasiri, N. & De, S. Wearable and miniaturized sensor technologies for personalized and
preventive medicine. _Adv. Funct. Mater._ 27, 1605271 (2017). Article Google Scholar * Yetisen, A. K., Martinez-Hurtado, J. L., Ünal, B., Khademhosseini, A. & Butt, H. Wearables in
medicine. _Adv. Mater._ 30, 1706910 (2018). Article PubMed PubMed Central Google Scholar * Kim, J., Campbell, A. S., de Ávila, B. E.-F. & Wang, J. Wearable biosensors for healthcare
monitoring. _Nat. Biotechnol._ 37, 389–406 (2019). Article CAS PubMed PubMed Central Google Scholar * Schalkamp, A.-K., Peall, K. J., Harrison, N. A. & Sandor, C. Wearable
movement-tracking data identify Parkinson’s disease years before clinical diagnosis. _Nat. Med._ 29, 2048–2056 (2023). Article CAS PubMed Google Scholar * Wang, J., Wu, B., Wei, P., Sun,
S. & Wu, P. Fatigue-free artificial ionic skin toughened by self-healable elastic nanomesh. _Nat. Commun._ 13, 4411 (2022). Article ADS CAS PubMed PubMed Central Google Scholar *
Tang, L., Shang, J. & Jiang, X. Multilayered electronic transfer tattoo that can enable the crease amplification effect. _Sci. Adv._ 7, eabe3778 (2021). Article ADS CAS PubMed PubMed
Central Google Scholar * Wang, Y. et al. Electrically compensated, tattoo-like electrodes for epidermal electrophysiology at scale. _Sci. Adv._ 6, eabd0996 (2020). Article ADS CAS
PubMed PubMed Central Google Scholar * He, W. et al. Integrated textile sensor patch for real-time and multiplex sweat analysis. _Sci. Adv._ 5, eaax0649 (2019). Article ADS CAS PubMed
PubMed Central Google Scholar * Yin, L. et al. A self-sustainable wearable multi-modular E-textile bioenergy microgrid system. _Nat. Commun._ 12, 1542 (2021). Article ADS CAS PubMed
PubMed Central Google Scholar * Hong, Y. J., Jeong, H., Cho, K. W., Lu, N. & Kim, D. H. Wearable and implantable devices for cardiovascular healthcare: from monitoring to therapy based
on flexible and stretchable electronics. _Adv. Funct. Mater._ 29, 1808247 (2019). Article Google Scholar * Li, M. et al. Flexible accelerated‐wound‐healing antibacterial MXene‐based
epidermic sensor for intelligent wearable human‐machine interaction. _Adv. Funct. Mater._ 32, 2208141 (2022). Article CAS Google Scholar * Zhang, Q. et al. Wearable triboelectric sensors
enabled gait analysis and waist motion capture for IoT‐based smart healthcare applications. _Adv. Sci._ 9, 2103694 (2022). Article CAS Google Scholar * Iqbal, S. M. A., Mahgoub, I., Du,
E., Leavitt, M. A. & Asghar, W. Advances in healthcare wearable devices. _npj Flex. Electron_ 5, 9 (2021). Article Google Scholar * Kim, H., Kim, E., Choi, C. & Yeo, W.-H. Advances
in soft and dry electrodes for wearable health monitoring devices. _Micromachines_ 13, 629 (2022). Article PubMed PubMed Central Google Scholar * Zhang, L. et al. Fully organic
compliant dry electrodes self-adhesive to skin for long-term motion-robust epidermal biopotential monitoring. _Nat. Commun._ 11, 4683 (2020). Article ADS CAS PubMed PubMed Central
Google Scholar * Shimura, T., Sato, S., Zalar, P. & Matsuhisa, N. Engineering the comfort-of-wear for next generation wearables. _Adv. Electron. Mater._ 9, 2200512 (2022). Article
Google Scholar * Zhang, J. et al. Scalable manufacturing of free-standing, strong Ti3C2Tx MXene films with outstanding conductivity. _Adv. Mater._ 32, 2001093 (2020). Article CAS Google
Scholar * Cao, F. et al. Recent advances in oxidation stable chemistry of 2D MXenes. _Adv. Mater._ 34, 2107554 (2022). Article CAS Google Scholar * Dong, H., Sun, J., Liu, X., Jiang, X.
& Lu, S. Highly sensitive and stretchable MXene/CNTs/TPU composite strain sensor with bilayer conductive structure for human motion detection. _ACS Appl. Mater. Interfaces_ 14,
15504–15516 (2022). Article CAS PubMed Google Scholar * Kazmierczak, N. P. et al. Strain fields in twisted bilayer graphene. _Nat. Mater._ 20, 956–963 (2021). Article ADS CAS PubMed
Google Scholar * Wu, N. et al. Ultrafine cellulose nanofiber-assisted physical and chemical cross-linking of mxene sheets for electromagnetic interference shielding. _Small Methods_ 5,
2100889 (2021). Article CAS Google Scholar * Park, G. S. et al. Comb-type polymer-hybridized MXene nanosheets dispersible in arbitrary polar, nonpolar, and ionic solvents. _Sci. Adv._ 8,
eabl5299 (2022). Article CAS PubMed PubMed Central Google Scholar * Jang, H. et al. Graphene e-tattoos for unobstructive ambulatory electrodermal activity sensing on the palm enabled by
heterogeneous serpentine ribbons. _Nat. Commun._ 13, 6604 (2022). Article ADS CAS PubMed PubMed Central Google Scholar * Christopher, R. N., Tim, A. S. & Elliott, J. R. Initial
design and experimental evaluation of a pneumatic interference actuator. _Soft Robot._ 5, 138–148 (2018). Article Google Scholar * Allan Joshua, V., Kyrian, S., & Herman, v.d.K. Soft,
wearable, and pleated pneumatic interference actuator provides knee extension torque for sit-to-stand. _Soft Robot._ 8, 28–43 (2021). Article Google Scholar * Hu, L. et al. Highly
conductive paper for energy-storage devices. _Proc. Natl Acad. Sci. USA_ 106, 21490–21494 (2009). Article ADS CAS PubMed PubMed Central Google Scholar * Shehata, N. et al. Stretchable
nanofibers of polyvinylidenefluoride (PVDF)/thermoplastic polyurethane (TPU) nanocomposite to support piezoelectric response via mechanical elasticity. _Sci. Rep._ 12, 8335 (2022). Article
ADS CAS PubMed PubMed Central Google Scholar * Lin, T. A., Lou, C.-W. & Lin, J.-H. The effects of thermoplastic polyurethane on the structure and mechanical properties of modified
polypropylene blends. _Appl. Sci._ 7, 1254 (2017). Article Google Scholar * Proto, A. et al. Measurements of generated energy/electrical quantities from locomotion activities using
piezoelectric wearable sensors for body motion energy harvesting. _Sensors_ 16, 524 (2016). Article ADS PubMed PubMed Central Google Scholar * Pal, A., Restrepo, V., Goswami, D. &
Martinez, R. V. Exploiting mechanical instabilities in soft robotics: control, sensing, and actuation. _Adv. Mater._ 33, 2006939 (2021). Article CAS Google Scholar * Liu, X., Wang, H.,
Li, Z. & Qin, L. Deep learning in ECG diagnosis: a review. _Knowl. -Based Syst._ 227, 107187 (2021). Article Google Scholar * Chen, L.-K. et al. Asian Working Group for Sarcopenia:
2019 consensus update on sarcopenia diagnosis and treatment. _J. Am. Med. Dir. Assoc._ 21, 300–307 (2020). Article PubMed Google Scholar * Sadikoglu, F., Kavalcioglu, C. & Dagman, B.
Electromyogram (EMG) signal detection, classification of EMG signals and diagnosis of neuropathy muscle disease. _Proc. Comput. Sci._ 120, 422–429 (2017). Article Google Scholar * Meigal,
A. et al. Non-linear EMG parameters for differential and early diagnostics of Parkinson’s disease. _Front. Neurol._ 4, 135 (2013). Article PubMed PubMed Central Google Scholar * Lin, D.
C., McGowan, C. P., Blum, K. P. & Ting, L. H. Yank: the time derivative of force is an important biomechanical variable in sensorimotor systems. _J. Exp. Biol._ 222, jeb180414 (2019).
Article PubMed PubMed Central Google Scholar * Yuen, A. C. Y. et al. Study of structure morphology and layer thickness of Ti3C2 MXene with Small-Angle Neutron Scattering (SANS). _Compos.
Part C Open Access_ 5, 100155 (2021). Article CAS Google Scholar * Liu, R. et al. MXene-coated air-permeable pressure-sensing fabric for smart wear. _ACS Appl. Mater. Interfaces_ 12,
46446–46454 (2020). Article CAS PubMed Google Scholar * Goyal, K., Borkholder, D. A. & Day, S. W. Dependence of skin-electrode contact impedance on material and skin hydration.
_Sensors_ 22, 8510 (2022). Article ADS CAS PubMed PubMed Central Google Scholar * Miyamoto, A. et al. Inflammation-free, gas-permeable, lightweight, stretchable on-skin electronics
with nanomeshes. _Nat. Nanotechnol._ 12, 907–913 (2017). Article ADS CAS PubMed Google Scholar * Liu, Y. et al. Epidermal mechano-acoustic sensing electronics for cardiovascular
diagnostics and human-machine interfaces. _Sci. Adv._ 2, e1601185 (2016). Article ADS PubMed PubMed Central Google Scholar * Cao, W.-T. et al. Binary strengthening and toughening of
MXene/cellulose nanofiber composite paper with nacre-inspired structure and superior electromagnetic interference shielding properties. _ACS Nano_ 12, 4583–4593 (2018). Article CAS PubMed
Google Scholar * Crow, R. M. & Dewar, M. M. Stresses in clothing as related to seam strength. _Text. Res. J._ 56, 467–473 (1986). Article Google Scholar * Han, M. & Shen, W.
Nacre-inspired cellulose nanofiber/MXene flexible composite film with mechanical robustness for humidity sensing. _Carbohydr. Polym._ 298, 120109 (2022). Article CAS PubMed Google Scholar
* Harvey, C. J., LeBouf, R. F. & Stefaniak, A. B. Formulation and stability of a novel artificial human sweat under conditions of storage and use. _Toxicol. Vitr._ 24, 1790–1796
(2010). Article CAS Google Scholar * Lee, H. et al. Stretchable array electromyography sensor with graph neural network for static and dynamic gestures recognition system. _npj Flex_.
_Electron_ 7, 20 (2023). Google Scholar * Ozdemir, M. A., Kisa, D. H., Guren, O. & Akan, A. Hand gesture classification using time–frequency images and transfer learning based on CNN.
_Biomed. Signal Process. Control_ 77, 103787 (2022). Article Google Scholar * Zhu, K., Guo, W., Yang, G., Li, Z. & Wu, H. High-fidelity recording of EMG signals by multichannel on-skin
electrode arrays from target muscles for effective human–machine interfaces. _ACS Appl. Electron. Mater._ 3, 1350–1358 (2021). Article CAS Google Scholar * Jiang, Y. Combination of
wearable sensors and internet of things and its application in sports rehabilitation. _Comput. Commun._ 150, 167–176 (2020). Article Google Scholar * Moin, A. et al. A wearable biosensing
system with in-sensor adaptive machine learning for hand gesture recognition. _Nat. Electron._ 4, 54–63 (2021). Article Google Scholar * Chen, L., Fu, J., Wu, Y., Li, H. & Zheng, B.
Hand gesture recognition using compact CNN via surface electromyography signals. _Sensors_ 20, 672 (2020). Article ADS PubMed PubMed Central Google Scholar * Gong Q., Jiang X., Liu Y.,
Yu M. & Hu Y. A flexible wireless sEMG system for wearable muscle strength and fatigue monitoring in real time. _Adv. Electron. Mater_. 9, 2200916 (2023). Download references
ACKNOWLEDGEMENTS This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korean government (MSIT) (No. RS-2023-00234581 and No. RS-2024-00415347) and
the Korea Medical Device Development Fund grant funded by the Korean government (the Ministry of Science and ICT, the Ministry of Trade, Industry and Energy, the Ministry of Health &
Welfare, and the Ministry of Food and Drug Safety) (Project Number: 1711174497, RS-2020-KD000093). AUTHOR INFORMATION Author notes * These authors contributed equally: Sanghyun Lee, Dong Hae
Ho, Janghwan Jekal. AUTHORS AND AFFILIATIONS * Department of Chemical and Biomolecular Engineering, Yonsei University, Seoul, Republic of Korea Sanghyun Lee, Soo Young Cho, Young Jin Choi
& Jeong Ho Cho * Department of Energy Science and Engineering, Daegu Gyeongbuk Institute of Science and Technology (DGIST), Daegu, Republic of Korea Dong Hae Ho * Department of Robotics
and Mechatronics Engineering, Daegu Gyeongbuk Institute of Science and Technology (DGIST), Daegu, Republic of Korea Janghwan Jekal, Saehyuck Oh & Kyung-In Jang * Department of Mechanical
Science and Engineering, University of Illinois at Urbana−Champaign, Urbana, IL, USA Yoon Young Choi * School of Electrical and Electronic Engineering, Yonsei University, Seoul, Republic of
Korea Taeyoon Lee * Department of Bio and Brain Engineering, Korea Institute of Science and Technology (KIST), Seoul, Republic of Korea Taeyoon Lee Authors * Sanghyun Lee View author
publications You can also search for this author inPubMed Google Scholar * Dong Hae Ho View author publications You can also search for this author inPubMed Google Scholar * Janghwan Jekal
View author publications You can also search for this author inPubMed Google Scholar * Soo Young Cho View author publications You can also search for this author inPubMed Google Scholar *
Young Jin Choi View author publications You can also search for this author inPubMed Google Scholar * Saehyuck Oh View author publications You can also search for this author inPubMed Google
Scholar * Yoon Young Choi View author publications You can also search for this author inPubMed Google Scholar * Taeyoon Lee View author publications You can also search for this author
inPubMed Google Scholar * Kyung-In Jang View author publications You can also search for this author inPubMed Google Scholar * Jeong Ho Cho View author publications You can also search for
this author inPubMed Google Scholar CONTRIBUTIONS Conceptualization: S.L. and D.H.H. Experiment: S.L., D.H.H., and J.J. Methodology: S.L., D.H.H., and J.J. Main data analysis and manuscript
writing: S.L., D.H.H., and J.J. Assisted data analysis and manuscript writing: S.Y.C., Y.J.C., S.O., and Y.Y.C. Visualization: S.L., D.H.H., and J.J. Supervision: T.L., K.-I.J., and J.H.C.
All authors contributed to the writing of the manuscript. All authors have approved the final version of the manuscript. CORRESPONDING AUTHORS Correspondence to Kyung-In Jang or Jeong Ho
Cho. ETHICS DECLARATIONS COMPETING INTERESTS The authors declare no competing interests. ADDITIONAL INFORMATION PUBLISHER’S NOTE Springer Nature remains neutral with regard to jurisdictional
claims in published maps and institutional affiliations. SUPPLEMENTARY INFORMATION SUPPLEMENTARY INFORMATION PEER REVIEW FILE DESCRIPTION OF ADDITIONAL SUPPLEMENTARY FILES SUPPLEMENTARY
MOVIE1 SUPPLEMENTARY MOVIE2 SUPPLEMENTARY MOVIE3 SUPPLEMENTARY MOVIE4 SUPPLEMENTARY MOVIE5 SUPPLEMENTARY MOVIE6 REPORTING SUMMARY SOURCE DATA SOURCE DATA RIGHTS AND PERMISSIONS OPEN ACCESS
This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as
long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third
party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the
article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright
holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. Reprints and permissions ABOUT THIS ARTICLE CITE THIS ARTICLE Lee, S., Ho, D.H., Jekal, J. _et
al._ Fabric-based lamina emergent MXene-based electrode for electrophysiological monitoring. _Nat Commun_ 15, 5974 (2024). https://doi.org/10.1038/s41467-024-49939-x Download citation *
Received: 10 October 2023 * Accepted: 20 June 2024 * Published: 02 October 2024 * DOI: https://doi.org/10.1038/s41467-024-49939-x SHARE THIS ARTICLE Anyone you share the following link with
will be able to read this content: Get shareable link Sorry, a shareable link is not currently available for this article. Copy to clipboard Provided by the Springer Nature SharedIt
content-sharing initiative